USE OF OPEN-ENDED COAXIAL LINE SENSOR WITH A LAMINAR OR LIQUID DIELECTRIC BACKED BY A CONDUCTING PLANE

被引:17
作者
JENKINS, S
WARHAM, AGP
CLARKE, RN
机构
[1] Natl Physical Lab, Teddington
关键词
COAXIAL LINE SENSOR; LIQUID DIELECTRIC;
D O I
10.1049/ip-h-2.1992.0033
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
A numerical technique has been developed which enables the open-ended coaxial line sensor to be used for measurements on laminar dielectric specimens which are backed by a conducting plane. The theory, which allows the reflection coefficient of the sensor to be computed from the complex permittivity of the specimen, has been tested using an automatic network analyser for measurements on reference liquids at frequencies between 10 MHz and 3 GHz. The discrepancies between theory and measurement do not exceed the estimated measurement uncertainties, provided that the specimen is sufficiently lossy to render reflections from its perimeter negligible. The ability to measure laminar specimens is an important extension to the applications of the sensor in dielectric metrology.
引用
收藏
页码:179 / 182
页数:4
相关论文
共 11 条
[1]   ANALYSIS OF AN OPEN-ENDED COAXIAL LINE SENSOR IN LAYERED DIELECTRICS [J].
ANDERSON, LS ;
GAJDA, GB ;
STUCHLY, SS .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1986, 35 (01) :13-18
[2]   A CRITICAL-STUDY OF THE OPEN-ENDED COAXIAL LINE SENSOR TECHNIQUE FOR RF AND MICROWAVE COMPLEX PERMITTIVITY MEASUREMENTS [J].
GRANT, JP ;
CLARKE, RN ;
SYMM, GT ;
SPYROU, NM .
JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1989, 22 (09) :757-770
[3]   DIELECTRIC MEASUREMENTS ON REFERENCE LIQUIDS USING AUTOMATIC NETWORK ANALYZERS AND CALCULABLE GEOMETRIES [J].
JENKINS, S ;
HODGETTS, TE ;
CLARKE, RN ;
PREECE, AW .
MEASUREMENT SCIENCE AND TECHNOLOGY, 1990, 1 (08) :691-702
[4]  
JENKINS S, 1991, UNPUB COMPLEX PERMIT
[6]   THE DIELECTRIC-PROPERTIES OF WATER AT MICROWAVE-FREQUENCIES [J].
KAATZE, U ;
UHLENDORF, V .
ZEITSCHRIFT FUR PHYSIKALISCHE CHEMIE-WIESBADEN, 1981, 126 (02) :151-165
[7]   NONINVASIVE ELECTRICAL CHARACTERIZATION OF MATERIALS AT MICROWAVE-FREQUENCIES USING AN OPEN-ENDED COAXIAL LINE - TEST OF AN IMPROVED CALIBRATION TECHNIQUE [J].
MISRA, D ;
CHABBRA, M ;
EPSTEIN, BR ;
MIROTZNIK, M ;
FOSTER, KR .
IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 1990, 38 (01) :8-14
[8]   THE ANNULAR SLOT ANTENNA IN A LOSSY BIOLOGICAL MEDIUM [J].
NEVELS, RD ;
BUTLER, CM ;
YABLON, W .
IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 1985, 33 (04) :314-319
[9]  
Nishikata A., 1988, Electronics and Communications in Japan, Part 2 (Electronics), V71, P95, DOI 10.1002/ecjb.4420710610
[10]  
WARHAM AGP, 1989, DITC15289 NAT PHYS L