CMOS CHIP FOR INVASIVE ULTRASOUND IMAGING

被引:32
作者
BLACK, WC [1 ]
STEPHENS, DN [1 ]
机构
[1] ENDOSON CORP,IMAGING TECHNOL,RANCHO CORDOVA,CA 95670
关键词
D O I
10.1109/4.328640
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A very small transmit/receiver chip has been developed for use in an arterial ultrasonic imaging system. In this technique, a solid-state ultrasonic imaging head placed within a small medical catheter is used to provide high quality 360 degrees images of arteries as small as 2 mm in diameter. Novel design and packaging techniques have been used to allow four easily testable 0.86 mm x 1.65 mm mixed-signal CMOS die to be placed on a multichip carrier within this 1.83 mm diameter imaging probe. Each chip contains interface circuitry for sixteen transducers including 20 MHz transmit pulsers and receive current amplifiers with approximately 1.3 pA/rt-Hz equivalent input noise performance. The techniques described here are generally applicable to any probe or device with extreme size and performance requirements.
引用
收藏
页码:1381 / 1387
页数:7
相关论文