CONVENTIONAL CHEST RADIOGRAPHY VS DUAL-ENERGY COMPUTED RADIOGRAPHY IN THE DETECTION AND CHARACTERIZATION OF PULMONARY NODULES

被引:75
作者
KELCZ, F
ZINK, FE
PEPPLER, WW
KRUGER, DG
ERGUN, DL
MISTRETTA, CA
机构
[1] UNIV WISCONSIN,DEPT MED PHYS,MADISON,WI 53706
[2] PHILIPS MED SYST NA,SHELTON,CT 06484
关键词
D O I
10.2214/ajr.162.2.8310908
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. We evaluated a single-exposure, phosphor-plate, dual-energy imaging device that produces, in addition to conventional chest radiographs, both tissue- and bone-selective images. Our purpose was to determine whether dual-energy radiography was more accurate than routine chest radiography for detection and characterization of pulmonary nodules. SUBJECTS AND METHODS. Two hundred patients undergoing chest CT were asked to volunteer to have dual-energy and conventional chest radiographs obtained immediately before or after their CT scan. Radiographs from a subset of 50 of these patients with 116 CT-detected nodules and 10 patients with normal findings on CT scans of the chest were presented to the observers for the nodule detection study. Similarly, radiographs from a subset of 29 patients with 20 calcified and 20 uncalcified nodules were presented to five observers to determine nodule calcification. Dual-energy images were produced by filtering the X-ray tube output with a gadolinium sheet while using a multiple phosphor plate receptor. A dual-energy triad of images consisting of a conventional image, a tissue-selective image, and a bone-selective image were produced. The conventional chest radiographs and dual-energy image sets were presented to observers in random order. Data from a free response receiver operating curve and a receiver operating curve were generated for nodule detection and characterization, respectively. RESULTS. By using the dual-energy images, all five observers improved their ability to diagnose pulmonary nodules (p = .0005) and to characterize nodules as calcified (p = .005). CONCLUSION. By eliminating rib shadows with tissue-selective images and enhancing calcified structures with bone-selective images, dual-energy chest radiography improved the ability of all observers, regardless of expertise, to detect and characterize pulmonary nodules.
引用
收藏
页码:271 / 278
页数:8
相关论文
共 33 条
[1]   STRATEGIES FOR IMPROVING POWER IN DIAGNOSTIC-RADIOLOGY RESEARCH [J].
BEAM, CA .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1992, 159 (03) :631-637
[2]  
BERKSON J, 1960, AM REV RESPIR DIS, V81, P660
[3]   DETECTION OF BULLOUS LUNG-DISEASE - CONVENTIONAL RADIOGRAPHY VS DIGITAL STORAGE PHOSPHOR RADIOGRAPHY [J].
BUCKLEY, KM ;
SCHAEFER, CM ;
GREENE, R ;
AGATSTON, S ;
FAY, J ;
LLEWELLYN, HJ ;
MROSE, HE ;
RUBENS, JR .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1991, 156 (03) :467-470
[4]   FREE-RESPONSE METHODOLOGY - ALTERNATE ANALYSIS AND A NEW OBSERVER-PERFORMANCE EXPERIMENT [J].
CHAKRABORTY, DP ;
WINTER, LHL .
RADIOLOGY, 1990, 174 (03) :873-881
[5]   SINGLE-EXPOSURE DUAL-ENERGY COMPUTED RADIOGRAPHY - IMPROVED DETECTION AND PROCESSING [J].
ERGUN, DL ;
MISTRETTA, CA ;
BROWN, DE ;
BYSTRIANYK, RT ;
SZE, WK ;
KELCZ, F ;
NAIDICH, DP .
RADIOLOGY, 1990, 174 (01) :243-249
[6]   CALCIFICATION IN PULMONARY NODULES - DETECTION WITH DUAL-ENERGY DIGITAL RADIOGRAPHY [J].
FRASER, RG ;
HICKEY, NM ;
NIKLASON, LT ;
SABBAGH, EA ;
LUNA, RF ;
ALEXANDER, CB ;
ROBINSON, CA ;
KATZENSTEIN, ALA ;
BARNES, GT .
RADIOLOGY, 1986, 160 (03) :595-601
[7]  
Glocker R, 1925, ANN PHYS-BERLIN, V76, P369
[8]   MISSED LUNG NODULES - LOST OPPORTUNITIES FOR CANCER CURE [J].
GREENE, RE .
RADIOLOGY, 1992, 182 (01) :8-9
[9]   COMPARISON OF DUAL-ENERGY AND CONVENTIONAL CHEST RADIOGRAPHY FOR NODULE DETECTION [J].
HO, JT ;
KRUGER, RA .
INVESTIGATIVE RADIOLOGY, 1989, 24 (11) :861-868
[10]  
IM JG, 1980, AJR, V135, P1