ULTRAVIOLET RAMAN-SPECTROSCOPY USING AN ATOMIC VAPOR FILTER AND INCOHERENT EXCITATION

被引:15
作者
PELLETIER, MJ
机构
[1] The Procter & Gamble Company, Miami Valley Laboratories, P.O. Box 398707, Cincinnati,OH,45239-8707, United States
关键词
RAMAN; ULTRAVIOLET; MERCURY; LINE-REJECTION FILTER; ANTI-STOKES;
D O I
10.1366/0003702924125320
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A far-UV line-rejection filter for Raman spectroscopy having an optical density greater than 5.5, a rejection bandwidth of only a few wavenumbers, and a transmission greater than 80% throughout the Stokes and anti-Stokes regions of the Raman spectrum is described. Raman spectra were collected with the use of this filter to remove elastically scattered light. Mercury penlamps were used for excitation. Good-quality spectra could be obtained with integration times as short as 1 second per wave-number. The advantages of using an atomic line source rather than a laser-based source to excite far-UV Raman spectra are discussed.
引用
收藏
页码:395 / 400
页数:6
相关论文
共 52 条