PERFORMANCE-CHARACTERISTICS OF AN 8-RING WHOLE-BODY PET SCANNER

被引:206
作者
KOPS, ER [1 ]
HERZOG, H [1 ]
SCHMID, A [1 ]
HOLTE, S [1 ]
FEINENDEGEN, LE [1 ]
机构
[1] KAROLINSKA HOSP INST,DEPT NEURORADIOL,STOCKHOLM,SWEDEN
关键词
Attenuation Values; Emission computed tomography; Performance Evaluation; Techniques;
D O I
10.1097/00004728-199005000-00022
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The technical characteristics of the multislice whole-body positron emission tomographic scanner (model PC4096-15WB Scanditronix) and its performance parameters are described. Spatial resolution at the center of the field of view was found to be 4.9 mm in-plane and 4.6 mm (cross slices) and 6.0 mm (direct slices) in the axial direction. The sensitivity for true and scattered coincidences is ~5,000 cps for direct slices and 7,100 cps for cross slices. At an activity concentration of 37 kBq/ml the system deadtime was ~5%. By measuring a uniform phantom with a cold cylindrical insert (5.0 cm diameter), the scatter fraction was found to be ~5%. The mean global uniformity over all 15 slices was 6.5%, whereas the local uniformity was found to be 4.3%. No systematic nonuniformities were observed. Finally, various methods for attenuation correction (transmission scan, contour finding, ellipse) were utilized to test their effects on the resulting reconstructed images. © 1990 Raven Press, Ltd., New York.
引用
收藏
页码:437 / 445
页数:9
相关论文
共 13 条
  • [1] CORRECTION FOR SCATTERED RADIATION IN A RING DETECTOR POSITRON CAMERA BY INTEGRAL TRANSFORMATION OF THE PROJECTIONS
    BERGSTROM, M
    ERIKSSON, L
    BOHM, C
    BLOMQVIST, G
    LITTON, J
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1983, 7 (01) : 42 - 50
  • [2] HUMAN CEREBRAL GLUCOSE-METABOLISM DETERMINED BY POSITRON EMISSION TOMOGRAPHY - A REVISIT
    BROOKS, RA
    HATAZAWA, J
    DICHIRO, G
    LARSON, SM
    FISHBEIN, DS
    [J]. JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 1987, 7 (04) : 427 - 432
  • [3] CARSON RE, 1988, J NUCL MED, V29, P1558
  • [4] FIGURES OF MERIT FOR DIFFERENT DETECTOR CONFIGURATIONS UTILIZED IN HIGH-RESOLUTION POSITRON CAMERAS
    ERIKSSON, L
    BERGSTROM, M
    BOHM, C
    HOLTE, S
    KESSELBERG, M
    LITTON, J
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1986, 33 (01) : 446 - 451
  • [5] COMPARATIVE REGIONAL-ANALYSIS OF 2-FLUORODEOXYGLUCOSE AND METHYLGLUCOSE UPTAKE IN BRAIN OF 4 STROKE PATIENTS - WITH SPECIAL REFERENCE TO THE REGIONAL ESTIMATION OF THE LUMPED CONSTANT
    GJEDDE, A
    WIENHARD, K
    HEISS, WD
    KLOSTER, G
    DIEMER, NH
    HERHOLZ, K
    PAWLIK, G
    [J]. JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 1985, 5 (02) : 163 - 178
  • [6] REGIONAL KINETIC CONSTANTS AND CEREBRAL METABOLIC-RATE FOR GLUCOSE IN NORMAL HUMAN VOLUNTEERS DETERMINED BY DYNAMIC POSITRON EMISSION TOMOGRAPHY OF [F-18]-2-FLUORO-2-DEOXY-D-GLUCOSE
    HEISS, WD
    PAWLIK, G
    HERHOLZ, K
    WAGNER, R
    GOLDNER, H
    WIENHARD, K
    [J]. JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 1984, 4 (02) : 212 - 223
  • [7] A PRELIMINARY EVALUATION OF A POSITRON CAMERA SYSTEM USING WEIGHTED DECODING OF INDIVIDUAL CRYSTALS
    HOLTE, S
    ERIKSSON, L
    LARSSON, JE
    ERICSON, T
    STJERNBERG, H
    HANSEN, P
    BOHM, C
    KESSELBERG, M
    ROTA, E
    HERZOG, H
    FEINENDEGEN, L
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1988, 35 (01) : 730 - 734
  • [8] A PRELIMINARY EVALUATION OF A DUAL CRYSTAL POSITRON CAMERA
    HOLTE, S
    OSTERTAG, H
    KESSELBERG, M
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1987, 11 (04) : 691 - 697
  • [9] A QUAD BGO DETECTOR AND ITS TIMING AND POSITIONING DISCRIMINATION FOR POSITRON COMPUTED-TOMOGRAPHY
    MURAYAMA, H
    NOHARA, N
    TANAKA, E
    HAYASHI, T
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1982, 192 (2-3): : 501 - 511
  • [10] NUTT R, 1985, J NUCL MED, V26, pP28