FAST METHODS FOR INCLUDING ATTENUATION IN THE EM ALGORITHM

被引:45
作者
HEBERT, TJ [1 ]
LEAHY, R [1 ]
机构
[1] UNIV SO CALIF,INST SIGNAL & IMAGE PROC,DEPT ELECT ENGN,LOS ANGELES,CA 90089
关键词
D O I
10.1109/23.106710
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Attenuation of γ-rays within the body has for some time been recognized as one of the major factors affecting emission tomography data. The manner in which it is treated impacts on the accuracy of the reconstruction. An exact treatment of known or estimated attenuation can be formulated for least squares or maximum likelihood solutions by direct inclusion of the attenuation probabilities in the forward/back projections found in typical iterative algorithms. This approach has required ray tracing and summing of the attenuation coefficients along each ray path which is extremely computationally cumbersome. A fast approach to including attenuation in iterative maximum likelihood and least squares algorithms for SPECT is presented. Ray-tracing and summing of attenuation coefficients are replaced by the use of two look-up tables, one to compute attenuated ray path integrals based on a set of polar grid points and one to perform polar-to-rectangular transformations. The resulting algorithm implements a spatial average which is comparable in accuracy to ray-tracing with rectangular pixels yet requires less than one sixteenth the CPU time. Copyright © 1990 by The Institute of Electrical and Electronics Engineers, Inc.
引用
收藏
页码:754 / 758
页数:5
相关论文
共 16 条
[1]  
Barrett H. H., 1981, RADIOLOGICAL IMAGING
[2]  
BARRETT HH, 1981, RADIOLOGICAL IMAGING, V2
[3]   COMPENSATION OF TISSUE ABSORPTION IN EMISSION TOMOGRAPHY [J].
BELLINI, S ;
PIACENTINI, M ;
CAFFORIO, C ;
ROCCA, F .
IEEE TRANSACTIONS ON ACOUSTICS SPEECH AND SIGNAL PROCESSING, 1979, 27 (03) :213-218
[4]  
CHANG L, 1979, IEEE T NUC SCI, V26
[5]  
DEMPSTER AP, 1977, J ROYAL STATISTICA B, V39
[6]   AN INVESTIGATION OF THE STATIONARITY OF THE 3D MODULATION TRANSFER-FUNCTION OF SPECT [J].
GLICK, SJ ;
KING, MA ;
KNESAUREK, K ;
BURBANK, K .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1989, 36 (01) :973-977
[7]   AN ATTENUATED PROJECTOR BACKPROJECTOR FOR ITERATIVE SPECT RECONSTRUCTION [J].
GULLBERG, GT ;
HUESMAN, RH ;
MALKO, JA ;
PELC, NJ ;
BUDINGER, TF .
PHYSICS IN MEDICINE AND BIOLOGY, 1985, 30 (08) :799-816
[8]   A GENERALIZED EM ALGORITHM FOR 3-D BAYESIAN RECONSTRUCTION FROM POISSON DATA USING GIBBS PRIORS [J].
HEBERT, T ;
LEAHY, R .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1989, 8 (02) :194-202
[9]   FAST MLE FOR SPECT USING AN INTERMEDIATE POLAR REPRESENTATION AND A STOPPING CRITERION [J].
HEBERT, T ;
LEAHY, R ;
SINGH, M .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1988, 35 (01) :615-619