BEAM STEERING WITH PULSED 2-DIMENSIONAL TRANSDUCER ARRAYS

被引:220
作者
TURNBULL, DH
FOSTER, FS
机构
[1] Sunnybrook Health Science Centre, Reichmann Research Building
[2] Department of Medical Biophysics, University of Toronto, Toronto, ON
基金
英国医学研究理事会;
关键词
D O I
10.1109/58.84270
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Two-dimensional (2-D) transducer arrays offer the potential for producing symmetrically focused ultrasound beams that can be steered throughout a three-dimensional (3-D) volume. The major problem facing the development of 2-D arrays for imaging is the complexity arising from the large number of elements anticipated in such transducers. We have undertaken a theoretical investigation of the focusing and steering properties of pulsed 2-D arrays to characterize the parameters required for medical imaging, such as element size, spacing and number of elements. Details of the computational methods employed are presented in this paper, as well as a discussion of the steered beam properties of wideband 2-D arrays. The effects of apodization and element cross-coupling on the beam properties of a 2-D transducer array are examined. Finally, we discuss the beam properties of various sparse arrays having elements randomly distributed over the aperture of the transducer.
引用
收藏
页码:320 / 333
页数:14
相关论文
共 24 条
  • [1] TRANSIENT FIELDS OF CONCAVE ANNULAR ARRAYS
    ARDITI, M
    FOSTER, FS
    HUNT, JW
    [J]. ULTRASONIC IMAGING, 1981, 3 (01) : 37 - 61
  • [2] CHESTON TC, 1970, RADAR HDB, pCH11
  • [3] ANOMALOUS BEHAVIOR IN THE RADIATION-PATTERN OF PIEZOELECTRIC TRANSDUCERS INDUCED BY PARASITIC LAMB WAVE GENERATION
    DELANNOY, B
    BRUNEEL, C
    HAINE, F
    TORGUET, R
    [J]. JOURNAL OF APPLIED PHYSICS, 1980, 51 (07) : 3942 - 3948
  • [4] INFINITE PLANAR BAFFLES PROBLEM IN ACOUSTIC RADIATION AND ITS EXPERIMENTAL-VERIFICATION
    DELANNOY, B
    LASOTA, H
    BRUNEEL, C
    TORGUET, R
    BRIDOUX, E
    [J]. JOURNAL OF APPLIED PHYSICS, 1979, 50 (08) : 5189 - 5195
  • [5] DEVELOPMENT OF A 12 ELEMENT ANNULAR ARRAY TRANSDUCER FOR REALTIME ULTRASOUND IMAGING
    FOSTER, FS
    LARSON, JD
    MASON, MK
    SHOUP, TS
    NELSON, G
    YOSHIDA, H
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 1989, 15 (07) : 649 - 659
  • [6] GELLY JF, 1981, 1981 P IEEE ULTR S, P685
  • [7] Goodman J., 2005, INTRO FOURIER OPTICS
  • [8] ULTRASOUND TRANSDUCERS FOR PULSE-ECHO MEDICAL IMAGING
    HUNT, JW
    ARDITI, M
    FOSTER, FS
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1983, 30 (08) : 453 - 481
  • [9] DESIGN OF SLOTTED TRANSDUCER ARRAYS WITH MATCHED BACKINGS
    KINO, GS
    DESILETS, CS
    [J]. ULTRASONIC IMAGING, 1979, 1 (03) : 189 - 209
  • [10] KOJIMA T, 1986, 1986 P IEEE ULTR S, P649