IMPLEMENTATION OF SIMULTANEOUS ATTENUATION AND DETECTOR RESPONSE CORRECTION IN SPECT

被引:182
作者
TSUI, BMW
HU, HB
GILLAND, DR
GULLBERG, GT
机构
[1] UNIV UTAH,MED CTR,DEPT RADIOL,SALT LAKE CITY,UT 84132
[2] UNIV N CAROLINA,CURRICULUM BIOMED ENGN,CHAPEL HILL,NC 27514
关键词
BIOMEDICAL ENGINEERING - Computerized Tomography - IMAGE PROCESSING - Reconstruction - PHOTONS - Attenuation - Signal filtering and prediction;
D O I
10.1109/23.12831
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Simultaneous correction of nonuniform attenuation and detector response was implemented in single-photon-emission computed tomography (SPECT) image reconstruction. A ray-driven projector-backprojector that exactly models attenuation in the reconstructed image slice and the spatially variant detector response was developed and used in the iterative maximum-likelihood algorithm for the correction. A computer-generated heart-lung phantom was used in simulation studies to compare the simultaneous correction method with an intrinsic attenuation correction method using a smoothing filter, and intrinsic attenuation correction method using a deconvolution filter, and a modified Chang attenuation correction method using a nonuniform attenuation distribution. The results demonstrate that the present method provides more-accurate quantitation and superior image quality.
引用
收藏
页码:778 / 783
页数:6
相关论文
共 24 条
[1]  
BAILEY DL, 1987, J NUCL MED, V28, P844
[2]  
Budinger T. F., 1979, Image reconstruction from projections. Implementation and applications, P147
[3]   3-DIMENSIONAL RECONSTRUCTION IN NUCLEAR-MEDICINE EMISSION IMAGING [J].
BUDINGER, TF ;
GULLBERG, GT .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1974, NS21 (03) :2-20
[5]  
FLOYD CE, 1986, J NUCL MED, V27, P1577
[6]   AN ATTENUATED PROJECTOR BACKPROJECTOR FOR ITERATIVE SPECT RECONSTRUCTION [J].
GULLBERG, GT ;
HUESMAN, RH ;
MALKO, JA ;
PELC, NJ ;
BUDINGER, TF .
PHYSICS IN MEDICINE AND BIOLOGY, 1985, 30 (08) :799-816
[7]   THE USE OF FILTERING METHODS TO COMPENSATE FOR CONSTANT ATTENUATION IN SINGLE-PHOTON EMISSION COMPUTED-TOMOGRAPHY [J].
GULLBERG, GT ;
BUDINGER, TF .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1981, 28 (02) :142-157
[8]   METHODS OF 3-DIMENSIONAL RECONSTRUCTION FROM A SET OF RADIOISOTOPE SCINTIGRAMS [J].
HSIEH, RC ;
WEE, WG .
IEEE TRANSACTIONS ON SYSTEMS MAN AND CYBERNETICS, 1976, 6 (12) :854-862
[9]  
KAY DB, 1975, J NUCL MED, V16, P540
[10]  
KING MA, 1987, J NUCL MED, V28, P1192