PERFORMANCE EVALUATION OF CONTEMPORARY SPIROMETERS

被引:104
作者
NELSON, SB
GARDNER, RM
CRAPO, RO
JENSEN, RL
机构
[1] UNIV UTAH, SCH MED, DEPT MED, DIV PULM, SALT LAKE CITY, UT 84112 USA
[2] UNIV UTAH, SCH MED, DEPT INFORMAT & MED, SALT LAKE CITY, UT 84112 USA
关键词
D O I
10.1378/chest.97.2.288
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
A comprehensive evaluation of 62 spirometers from 37 different sources was performed using a two-part protocol: calibrated syringe, and dynamic waveform testing. All testing was done with ambient air. Calibrated syringe testing examined the ability of the spirometers to accurately measure the output of a 3 L calibrating syringe under varying conditions. The accuracy, FVC volume linearity, and stability of each spirometer was determined from these data. All but five of 42 spirometers accurately measured a 3 L calibrating syringe to within ± 3 percent. Dynamic waveform testing consisted of introducing 24 standard waveforms into the spirometer from a computed-controlled air pump. The values of FVC, FEV1, and FEF25-75% were compared to the actual values for each waveform to determine a performance rating. Only 35 (56.5 percent) of the spirometers performed acceptably when measuring the 24 standard waveforms. Nine (14.5 percent) were marginal and 18 (29.0 percent) were unacceptable. Fifty-nine (95 percent) of the 62 spirometers were computerized. Software errors were found in 25 percent of the computerized systems evaluated. Although using a 3 L syringe for quality control purposes is essential, simple testing of spirometers with a 3 L calibrating syringe for validation purposes was inadequate to assess spirometer performance when compared to dynamic waveform testing. Dynamic waveform testing is essential to accurately measure and validate acceptability of spirometer system performance.
引用
收藏
页码:288 / 297
页数:10
相关论文
共 22 条
  • [1] [Anonymous], 1979, AM REV RESPIR DIS, V119, P831
  • [2] CRAPO RO, 1987, AM REV RESPIR DIS, V135, P1221
  • [3] ENRIGHT PL, 1987, OFFICE SPIROMETRY
  • [4] EVALUATION OF ELECTRONIC SPIROMETERS
    FITZGERALD, MX
    SMITH, AA
    GAENSLER, EA
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 1973, 289 (24) : 1283 - 1288
  • [5] GARDNER RM, 1980, AM REV RESPIR DIS, V121, P73
  • [6] SPIROMETRY - WHAT PAPER SPEED
    GARDNER, RM
    CRAPO, RO
    BILLINGS, RG
    SHIGEOKA, JW
    HANKINSON, JL
    [J]. CHEST, 1983, 84 (02) : 161 - 165
  • [7] GLINDMEYER HW, 1982, J OCCUP ENVIRON MED, V24, P1003
  • [8] EVALUATION OF THE TURBINE POCKET SPIROMETER
    GUNAWARDENA, KA
    HOUSTON, K
    SMITH, AP
    [J]. THORAX, 1987, 42 (09) : 689 - 693
  • [9] HANKINSON JL, 1982, AM REV RESPIR DIS, V126, P362
  • [10] HANKINSON JL, 1983, J APPL PHYSL REPIR E, V44, P1353