CLINICAL-EXPERIENCE WITH THE HEWLETT-PACKARD M-1350A FETAL MONITOR - CORRELATION OF DOPPLER-DETECTED FETAL BODY MOVEMENTS WITH FETAL HEART-RATE PARAMETERS AND PERINATAL OUTCOME

被引:25
作者
DEVOE, L
BOEHM, F
PAUL, R
FRIGOLETTO, F
PENSO, C
GOLDENBERG, R
RAYBURN, W
SMITH, C
机构
[1] VANDERBILT UNIV,SCH MED,DEPT OBSTET & GYNECOL,NASHVILLE,TN 37212
[2] UNIV SO CALIF,SCH MED,DEPT OBSTET & GYNECOL,LOS ANGELES,CA 90033
[3] HARVARD UNIV,SCH MED,DEPT OBSTET & GYNECOL,BOSTON,MA
[4] UNIV ALABAMA,SCH MED,BIRMINGHAM,AL
[5] UNIV NEBRASKA,SCH MED,OMAHA,NE 68198
关键词
ACTOCARDIOTOCOGRAPH; FETAL HEART RATE; FETAL MOVEMENT; DOPPLER ULTRASONOGRAPHY;
D O I
10.1016/S0002-9378(94)70243-8
中图分类号
R71 [妇产科学];
学科分类号
100211 ;
摘要
OBJECTIVE: Our purpose was to correlate measures of Doppler-detected fetal movements with standard fetal heart rate parameters and perinatal outcomes. STUDY DESIGN: This prospective, multiinstitutional trial used the Hewlett-Packard M1350A monitor to record simultaneous fetal heart rate baseline, variability, accelerations, decelerations, and number of fetal movements, and duration and percent of total time. These data were compared at 10- and 30-minute intervals during nonstress tests and were correlated with fetal heart rate baseline parameters and maternally perceived fetal movements and with outcomes of infants delivered within 7 days of the last test. RESULTS: At six centers 1704 actocardiograms from 884 third-trimester patients were analyzed. Doppler-detected fetal movement counts, durations, and percent of total time correlated weakly with all baseline fetal heart rate parameters (all values < 0.20). All fetal movement parameters increased significantly in successive 10-minute blocks and in periods of increased or normal fetal heart rate variability compared with those with fetal heart rate variability. The sensitivity, specificity, and predictive values of the percent of total movement time were comparable to those of standard nonstress test parameters. The risk of poor perinatal outcomes after nonreactive nonstress tests was lower in cases with fetal movements than in those without. CONCLUSIONS: Doppler actocardiography may help to discriminate fetal states during antepartum testing. It may prevent inappropriate diagnosis of fetal compromise when the nonstress test is nonreactive or nonreassuring.
引用
收藏
页码:650 / 655
页数:6
相关论文
共 14 条
[1]  
BASER I, 1992, OBSTET GYNECOL, V80, P62
[2]  
BESINGER RE, 1989, OBSTET GYNECOL, V74, P277
[3]   FOETAL HEART FREQUENCY AND PERINATAL CONDITION OF FOETUS AND NEWBORN [J].
HAMMACHER, K ;
HUTER, KA ;
BOKELMANN, J ;
WERNERS, PH .
GYNAECOLOGIA, 1968, 166 (04) :349-+
[4]  
JOHNSON TRB, 1990, OBSTET GYNECOL, V76, P42
[5]  
MAEDA K, 1987, ZBL GYNAKOL, V109, P873
[6]  
MAEDA K, 1986, INTRAUTERINE LIFE MA, P177
[7]   CHARACTERIZATION OF FETAL BODY MOVEMENT RECORDED BY THE HEWLETT-PACKARD M-1350-A FETAL MONITOR [J].
MELENDEZ, TD ;
RAYBURN, WF ;
SMITH, CV .
AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 1992, 167 (03) :700-702
[8]   ARE THERE BEHAVIORAL STATES IN THE HUMAN-FETUS [J].
NIJHUIS, JG ;
PRECHTL, HFR ;
MARTIN, CB ;
BOTS, RSGM .
EARLY HUMAN DEVELOPMENT, 1982, 6 (02) :177-195
[9]  
PATRICK J, 1984, AM J OBSTET GYNECOL, V48, P35
[10]   MATERNAL PERCEPTION OF FETAL MOVEMENTS AND REAL-TIME ULTRASOUND FINDINGS [J].
SCHMIDT, W ;
CSEH, I ;
HARA, K ;
KUBLI, F .
JOURNAL OF PERINATAL MEDICINE, 1984, 12 (06) :313-318