DETERMINATION OF GROWTH VALUE THRESHOLDS FOR BACTEC PLUS AEROBIC BLOOD CULTURE VIALS

被引:4
作者
MCGOWAN, JE [1 ]
METCHOCK, BG [1 ]
机构
[1] EMORY UNIV,SCH MED,DEPT PATHOL & LAB MED,ATLANTA,GA 30335
关键词
D O I
10.1128/JCM.30.4.771-774.1992
中图分类号
Q93 [微生物学];
学科分类号
071005 ; 100705 ;
摘要
Growth value thresholds used to identify positive blood culture vials can be defined by users for each BACTEC NR-660 bacteremia detection instrument. Growth values were compared with the recovery of organisms from vials flagged as positive during the testing of 3.056 high-volume vials containing aerobic (BACTEC PLUS 26) medium over a 2-month period. Results showed that optimal threshold values for our use of these vials varied from those recommended by the manufacturer; if the thresholds defined from these data had been used during the study period, total vials flagged as positive from which no organisms were recovered (false alarms) would have been reduced from 181 (5.9/100 vials tested) to 71 (2.3/100 vials tested), with a minimal decrease in the identification of vials containing usual or occasional pathogens (hits). Adjustments of growth value thresholds by the individual user can make the use of BACTEC instruments more efficient by decreasing further processing of vials from which no organisms are recovered.
引用
收藏
页码:771 / 774
页数:4
相关论文
共 11 条
[1]   ASSESSMENT OF THE BACTEC NR660 BLOOD CULTURE SYSTEM FOR THE DETECTION OF BACTEREMIA IN YOUNG-CHILDREN [J].
ANDERSON, JD ;
TROMBLEY, C ;
CIMOLAI, N .
JOURNAL OF CLINICAL MICROBIOLOGY, 1989, 27 (04) :721-723
[2]   ROUTINE EVALUATION OF THE NONRADIOMETRIC BACTEC NR-660 SYSTEM [J].
COURCOL, RJ ;
FRUCHART, A ;
ROUSSELDELVALLEZ, M ;
MARTIN, GR .
JOURNAL OF CLINICAL MICROBIOLOGY, 1986, 24 (01) :26-29
[3]   INFLUENCE OF BLOOD CARBON-DIOXIDE PRESSURE ON GROWTH INDEX VALUES IN NON-RADIOMETRIC BACTEC VIALS [J].
COURCOL, RJ ;
DUROCHER, AV ;
ROBIN, H ;
MARTIN, GR .
EUROPEAN JOURNAL OF CLINICAL MICROBIOLOGY & INFECTIOUS DISEASES, 1989, 8 (05) :400-401
[4]   CLINICAL COMPARISON OF A NEW AUTOMATED INFRARED BLOOD CULTURE SYSTEM WITH THE BACTEC-460 SYSTEM [J].
JUNGKIND, D ;
MILLAN, J ;
ALLEN, S ;
DYKE, J ;
HILL, E .
JOURNAL OF CLINICAL MICROBIOLOGY, 1986, 23 (02) :262-266
[5]   RAPID DETECTION OF POSITIVE BLOOD CULTURES WITH THE BACTEC NR-660 DOES NOT REQUIRE 1ST-DAY SUBCULTURING [J].
LEVI, MH ;
GIALANELLA, P ;
MOTYL, MR ;
MCKITRICK, JC .
JOURNAL OF CLINICAL MICROBIOLOGY, 1988, 26 (11) :2262-2265
[6]   DETECTION OF POSITIVE BLOOD CULTURES BY THE BACTEC NR660 - THE CLINICAL IMPORTANCE OF 5 VERSUS 7 DAYS OF TESTING [J].
MASTERSON, KC ;
MCGOWAN, JE .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1988, 90 (01) :91-94
[7]  
MICKELL SS, 1991, COMMUNICATION
[8]   PERFORMANCE OF A BACTEC NONRADIOMETRIC MEDIUM FOR PEDIATRIC BLOOD CULTURES [J].
MORELLO, JA ;
MATUSHEK, SM ;
DUNNE, WM ;
HINDS, DB .
JOURNAL OF CLINICAL MICROBIOLOGY, 1991, 29 (02) :359-362
[9]  
Swets J. A., 1982, EVALUATION DIAGNOSTI
[10]  
VANELSACKERNIELE AMW, 1989, EUR J CLIN MICROBIOL, V8, P724