DETERMINATION OF TOTAL COPPER IN HEMODIALYSIS WATER BY DIFFERENTIAL-PULSE ANODIC-STRIPPING VOLTAMMETRY

被引:9
作者
TAHAN, JE [1 ]
MORONTA, AJ [1 ]
ROMERO, RA [1 ]
机构
[1] UNIV ZULIA,FAC EXPTL CIENCIAS,INSTRUMENTAC ANALIT LAB,MARACAIBO 4011,VENEZUELA
关键词
Copper; Haemodialysis water;
D O I
10.1016/S0003-2670(00)83348-6
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An anodic stripping voltammetric method was developed in order to determine copper in the water used to prepare haemodialysis solutions. The interference from organic matter was overcome by high-pressure bomb mineralization. The electrochemical results were compared with those obtained by using graphite furnace atomic absorption spectrometry and the correlation was excellent (r = 0.983, p < 0.001). The detection limit was 0.2 μg l-1 copper. © 1990.
引用
收藏
页码:449 / 451
页数:3
相关论文
共 8 条
[1]  
LYLE WH, 1976, LANCET, V1, P1324
[2]  
MENDOZA JA, 1988, 15TH FACSS M BOST
[3]  
MORONTO AJ, 1990, 7TH INT S TRAC EL MA, P191
[4]  
NAVARRO JA, 1989, TRACE ELEM MED, V6, P70
[5]  
PARRA OE, 1987, ATOM SPECTROSC, V8, P105
[6]   ZINC, COPPER AND MAGNESIUM KINETICS DURING DESFERRIOXAMINE TREATMENT IN UREMIC PATIENTS [J].
SEGOLONI, GP ;
CANAVESE, C ;
DAMICONE, M ;
LAMON, S ;
TALARICO, S ;
VERCELLONE, A .
NEPHRON, 1984, 36 (04) :278-279
[7]  
TAHAN JE, 1989, 40TH PITTSB C AN CHE
[8]   THE EFFECTS OF COPPER-ZINC AND COPPER-CADMIUM INTERMETALLIC COMPOUNDS IN DIFFERENT SYSTEMS USED FOR ANODIC-STRIPPING VOLTAMMETRY [J].
WISE, JA ;
ROSTON, DA ;
HEINEMAN, WR .