AN ALTERNATIVE LIMB LEAD SYSTEM FOR ELECTROCARDIOGRAPHS IN EMERGENCY PATIENTS

被引:23
作者
TAKUMA, K
HORI, S
SASAKI, J
SHINOZAWA, Y
YOSHIKAWA, T
HANDA, S
HORIKAWA, M
AIKAWA, N
机构
[1] Department of Emergency Medicine, Keio University, School of Medicine, Tokyo
[2] Department of Internal Medicine, Keio University, School of Medicine, Tokyo
[3] the Department of Biomedical Engineering, School of High-Technology for Human Welfare, Tokai University, Tokyo
关键词
ECG; LEAD SYSTEM; PREHOSPITAL CARE; EMERGENCY;
D O I
10.1016/0735-6757(95)90160-4
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
It is occasionally difficult to record the standard 12-lead electrocardiograph (EGG) in emergency patients. The aim of this study was to evaluate the influence on electrocardiographic wave form recordings of moving the location of electrodes from the standard limb lead position to the trunk. The participants were 10 normal subjects and 20 patients with heart disease. In the new lead system, the limb electrodes were placed on the anterior acromial region and the anterior superior iliac spine using adhesive electrodes. Conventional 12-lead ECGs were recorded by the standard and the new lead system simultaneously in the supine position. Wave form analysis was done by an automatic analysis program. Motion artifacts in the recordings were less in the new lead system. The a wave amplitude of the new lead system increased in leads II, III and aVF, and decreased in leads I and aVL. However, the amplitudes of each wave obtained by standard electrocardiography and the new lead system correlated well (y = 1.008x + 2.038, r = 0.99, n = 2,880). In 99.6% of all wave forms, the differences in amplitudes were within 5% of the values of standard recordings. The average of differences in the ST-segment was 2.6 +/- 11.4 mu V. The frontal plane QRS axis obtained try the new lead system showed a vertical shift of 7.8 +/- 8.5 degrees (y = 0.94911x + 10.346, r = 0.98, n = 30). The recording errors produced by the new lead system were within the permissible range of variation. The new lead system is a reasonable alternative for recording ECGs if application of the standard lead is difficult in an emergency. (C) 1995 by W.B. Saunders Company
引用
收藏
页码:514 / 517
页数:4
相关论文
共 14 条
[1]  
International Electrotechnical Commission, Draft standard specification for performance of single and multi-channel electrocardiographs, (1978)
[2]  
International Electrotechnical Commission, Draft performance requirements for recording and analyzing single-channel and multi-channel electrocardiographs version 0.2, (1993)
[3]  
Bailey, Berson, Garson, Et al., Recommendations for standardization and specifications in automated electrocardiography, Bandwidth and digital signal processing, 81, pp. 730-739, (1990)
[4]  
Mason, Likar, A new system of multiple-lead exercise electrocardiography, American Heart Journal, 71, pp. 169-205, (1966)
[5]  
Edenbrandt, Pahlm, Sornmo, An accurate exercise lead system for bicycle ergometer tests, Eur Heart J, 10, pp. 268-272, (1989)
[6]  
Fesmire, Smith, Continuous 12-lead electrocardiograph monitoring in the emergency department, Am J Emerg Med, 11, pp. 54-60, (1993)
[7]  
Krucoff, Wagner, Pope, Et al., The portable programmable microprocessor-driven real-time 12-lead electrocardiograph monitor: A preliminary report of a new device for the noninvasive detection of successful reperfusion or silent coronary reocclusion, Am J Cardiol, 65, pp. 143-148, (1990)
[8]  
Rautaharju, Prineas, Crow, Et al., The effect of modified limb electrocardiographic wave amplitudes, J Electrocardiol, 13, pp. 109-113, (1980)
[9]  
Papouchado, Walker, James, Et al., Fundamental differences between the standard 12-lead electrocardiograph and the modified (Manson-Likar) exercise lead system, Eur Heart J, 8, pp. 725-733, (1987)
[10]  
Kleiner, Nelson, Boland, The 12-lead electrocardiogram in exercise testing, Arch Intern Med, 138, pp. 1572-1573, (1978)