EVALUATION OF A PROTOTYPE FOR A REFERENCE PLATELET COUNTER

被引:6
作者
LEWIS, SM
ROWAN, RM
KUBOTA, F
机构
[1] WESTERN INFIRM & ASSOCIATED HOSP,GLASGOW G11 6NT,SCOTLAND
[2] TOA MED ELECTR,KOBE,JAPAN
关键词
D O I
10.1136/jcp.43.11.932
中图分类号
R36 [病理学];
学科分类号
100104 ;
摘要
A semiautomated single channel aperture-impedance particle counter was developed as a prototype for a reference platelet count for assigning values to reference preparations used in automated blood cell counts. The instrument is equipped with sheath flow and an aperture orifice of 50 μm in diameter and 60 μm in length to eliminate non-axial flow and minimise coincidence errors. Use of fixed volume and red blood cell:platelet ratios obviate dilution errors. The counter was assessed in accordance with the International Committee for Standardization in Haematology protocol for evaluation of automated blood cell counters. The counter provided a high level of linearity and precision, accurate coincidence correction, controlled volume, stability and negligible carryover.
引用
收藏
页码:932 / 936
页数:5
相关论文
共 10 条
[1]  
[Anonymous], 1987, CLIN LAB HAEMATOL, V9, P73
[2]   REVISED SCHEME FOR EVALUATION OF AUTOMATIC INSTRUMENTS FOR USE IN CLINICAL CHEMISTRY [J].
BROUGHTON, PM ;
GOWENLOCK, AH ;
MCCORMACK, JJ ;
NEILL, DW .
ANNALS OF CLINICAL BIOCHEMISTRY, 1974, 11 (NOV) :207-218
[3]   A RAPID AND ACCURATE DIFFERENTIAL CENTRIFUGATION METHOD FOR PLATELET COUNTS [J].
COUSINS, S ;
LEWIS, SM .
JOURNAL OF CLINICAL PATHOLOGY, 1982, 35 (01) :114-116
[4]  
GRONER W, 1982, ADV HAEMATOLOGICAL M, P74
[5]  
Lewis S M, 1979, Clin Lab Haematol, V1, P227
[6]  
LEWIS SM, 1981, CLIN LAB HAEMATOL, V3, P215
[7]   COINCIDENCE CORRECTION IN RED BLOOD-CELL COUNTING [J].
LEWIS, SM ;
ENGLAND, JM ;
KUBOTA, F .
PHYSICS IN MEDICINE AND BIOLOGY, 1989, 34 (09) :1239-1246
[8]  
LEWIS SM, 1989, LAB PRACT, V38, P68
[9]  
1984, CLINLAB HAEMATOL, V6, P69
[10]  
1988, CLIN LAB HAEMATOL, V10, P203