ACCELERATOR-BASED EPITHERMAL NEUTRON BEAM DESIGN FOR NEUTRON-CAPTURE THERAPY

被引:70
作者
YANCH, JC
ZHOU, XL
SHEFER, RE
KLINKOWSTEIN, RE
机构
[1] MIT,WHITAKER COLL HLTH SCI & TECHNOL,CAMBRIDGE,MA 02139
[2] SCI RES LAB INC,SOMERVILLE,MA 02143
关键词
EPITHERMAL NEUTRONS; ACCELERATOR; NCT; MCNP;
D O I
10.1118/1.596815
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 [临床医学]; 100207 [影像医学与核医学]; 1009 [特种医学];
摘要
Recent interest in the production of epithermal neutrons for use in boron neutron capture therapy (BNCT) has prompted an investigation into the feasibility of generating such neutrons with a high current proton accelerator. Energetic protons (2.5 MeV) on a Li-7 target produce a spectrum of neutrons with maximum energy of roughly 800 keV. A number of combinations of D2O moderator, lead reflector, Li-6 thermal neutron filtration, and D2O/Li-6 shielding will result in a useful epithermal flux of 1.6 X 10(8) n/s at the patient position. The neutron beam is capable of delivering 3000 RBE-cGy to a tumor at a depth-of 7.5 cm in a total treatment time of 60-93 min (depending on RBE values used and based on a 24-cm diameter X 19-cm length D2O moderator). Treatment of deeper tumors with therapeutic advantage would also be possible. Maximum advantage depths (RBE weighted) of 8.2-9.2 (again depending on RBE values and precise moderator configuration) are obtained in a right-circular cylindrical phantom composed of brain-equivalent material with an advantage ratio of 4.7-6.3. A tandem cascade accelerator (TCA), designed and constructed at Science Research Laboratory (SRL) in Somerville MA, can provide the required proton beam parameters for BNCT of deep-seated tumors. An optimized configuration of materials required to shift the accelerator neutron spectrum down to therapeutically useful energies has been designed using Monte Carlo simulation in the Whitaker College Biomedical Imaging and Computation Laboratory at MIT. Actual construction of the moderator/reflector assembly is currently underway.
引用
收藏
页码:709 / 721
页数:13
相关论文
共 22 条
[1]
BARTH RF, 1990, CANCER RES, V50, P1061
[2]
BARTH RF, 1990, SCI AM OCT, P100
[3]
Briesmeister, 1986, MCNP GEN MONTE CARLO
[4]
EXPLANATION OF CEREBRAL WHITE-GRAY CONTRAST IN COMPUTED-TOMOGRAPHY [J].
BROOKS, RA ;
DICHIRO, G ;
KELLER, MR .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1980, 4 (04) :489-491
[5]
BRUGGER RM, 1990, NEUTRON BEAM DESIGN
[6]
CASWELL RS, 1982, INT J APPL RADIAT IS, V33, P1227, DOI 10.1016/0020-708X(82)90246-0
[7]
CHOI JR, 1990, IN PRESS 4TH P INT S
[8]
CHOI JR, 1990, NEUTRON BEAM DESIGN
[9]
CLEMENT SD, 1990, NEUTRON BEAM DESIGN
[10]
COMPARISON OF MEASURED PARAMETERS FROM A 24-KEV AND A BROAD-SPECTRUM EPITHERMAL NEUTRON BEAM FOR NEUTRON-CAPTURE THERAPY - AN IDENTIFICATION OF CONSEQUENTIAL PARAMETERS [J].
FAIRCHILD, RG ;
SARAF, SK ;
KALEFEZRA, J ;
LASTER, BH .
MEDICAL PHYSICS, 1990, 17 (06) :1045-1052