PATENT FORAMEN OVALE - A CAUSE OF HYPOXEMIA IN PATIENTS ON LEFT-VENTRICULAR SUPPORT

被引:43
作者
BALDWIN, RT [1 ]
DUNCAN, JM [1 ]
FRAZIER, OH [1 ]
WILANSKY, S [1 ]
机构
[1] TEXAS HEART INST,CULLEN CARDIOVASC RES LABS,POB 20345,HOUSTON,TX 77225
关键词
D O I
10.1016/0003-4975(91)91230-S
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We describe 2 patients who experienced right-to-left shunting during support with a left ventricular assist device as a result of patent foramen ovale. In the first patient, the patent foramen ovale was not found until autopsy. In the second patient, fluctuations in continuous mixed venous oxygen saturations caused us to suspect a patent foramen ovale, which was confirmed with transesophageal contrast echocardiography. We promptly repaired the defect and this patient survived. When patients on assist devices experience unexplained arterial hypoxemia, a patent foramen ovale should be considered. If a patent foramen ovale is found, surgical correction should be done early.
引用
收藏
页码:865 / 867
页数:3
相关论文
共 4 条
[1]   INCIDENCE AND SIZE OF PATENT FORAMEN OVALE DURING THE 1ST 10 DECADES OF LIFE - AN AUTOPSY STUDY OF 965 NORMAL HEARTS [J].
HAGEN, PT ;
SCHOLZ, DG ;
EDWARDS, WD .
MAYO CLINIC PROCEEDINGS, 1984, 59 (01) :17-20
[2]  
KYO S, 1989, ASAIO T, V35, P772
[3]   PREVALENCE OF RIGHT-TO-LEFT ATRIAL SHUNTING IN A HEALTHY POPULATION - DETECTION BY VALSALVA MANEUVER CONTRAST ECHOCARDIOGRAPHY [J].
LYNCH, JJ ;
SCHUCHARD, GH ;
GROSS, CM ;
WANN, LS .
AMERICAN JOURNAL OF CARDIOLOGY, 1984, 53 (10) :1478-1480
[4]  
ZAHN R, 1989, CIRCULATION S2, V80, P340