DETERMINATION OF PHENOLS AT THE NG/1 LEVEL IN DRINKING AND RIVER WATERS BY LIQUID-CHROMATOGRAPHY WITH UV AND ELECTROCHEMICAL DETECTION

被引:68
作者
RUANA, J [1 ]
URBE, I [1 ]
BORRULL, F [1 ]
机构
[1] UNIV ROVIRA & VIRGILI,DEPT QUIM,E-4300 TARRAGONA,SPAIN
关键词
D O I
10.1016/0021-9673(93)83226-I
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Solid-liquid extraction of samples and liquid chromatography with UV and electrochemical detection with laboratory-made microcolumns were applied to the separation and identification of phenols and substituted phenols in waters. The compounds mainly studied were the eleven phenols considered as priority pollutants by the US Environmental Protection Agency. Chromatographic separation was carried out with several water-methanol isocratic mobile phases; use of the autoincrement mode of the electrochemical detector allowed the compounds in the samples to be confirmed. The chromatographic system worked automatically. The detection limits obtained with prior concentration of the samples were 40-600 ng/l, depending of the phenol. Electrochemical detection was used for the determination of phenols in river and drinking waters; phenols at the ng/l level was detected
引用
收藏
页码:217 / 226
页数:10
相关论文
共 16 条
  • [1] DETERMINATION OF PHENOLS BY HPLC DOWN TO PPT LEVELS
    BALDWIN, DA
    DEBOWSKI, JK
    [J]. CHROMATOGRAPHIA, 1988, 26 : 186 - 190
  • [2] CROMPTON TR, 1985, DETERMINATION ORG, V2, P149
  • [3] FRESENIUS W, 1988, WATER ANAL, P536
  • [4] GRETZFELDHUSGEN A, 1990, HPLC1259521548 HEWL
  • [5] HENNION M-C, 1991, Aqua (Oxford), V40, P35
  • [6] PRIORITY POLLUTANTS I-A PERSPECTIVE VIEW
    KEITH, LH
    TELLIARD, WA
    [J]. ENVIRONMENTAL SCIENCE & TECHNOLOGY, 1979, 13 (04) : 416 - 423
  • [7] SENSITIVE LIQUID-CHROMATOGRAPHIC DETERMINATION OF ALKYL-PHENOLS, NITRO-PHENOLS AND CHLOROPHENOLS BY PRECOLUMN DERIVATIZATION WITH DANSYL CHLORIDE, POSTCOLUMN PHOTOLYSIS AND PEROXYOXALATE CHEMILUMINESCENCE DETECTION
    KWAKMAN, PJM
    KAMMINGA, DA
    BRINKMAN, UAT
    DEJONG, GJ
    [J]. JOURNAL OF CHROMATOGRAPHY, 1991, 553 (1-2): : 345 - 356
  • [8] HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY OF 11 PRIORITY SUBSTITUTED PHENOLS BY ISOCRATIC ELUTION
    LEE, HK
    LI, SFY
    TAY, YH
    [J]. JOURNAL OF CHROMATOGRAPHY, 1988, 438 (02): : 429 - 432
  • [9] MONTGOMERY JM, 1985, WATER TREATMENT PRIN, P385
  • [10] ONG CP, 1989, J CHROMATOGR, V464, P405