ANALYTIC REPRESENTATION OF THE BACKSCATTER CORRECTION FACTOR AT THE EXIT OF HIGH-ENERGY PHOTON BEAMS

被引:8
作者
KAPPAS, K
ROSENWALD, JC
机构
[1] INST CURIE,UNITE RADIOPHYS,26 RUE ULM,F-75231 PARIS 05,FRANCE
[2] UNIV PATRAS,DEPT MED PHYS,PATRAS,GREECE
关键词
DOSIMETRY; BACKSCATTER CORRECTION; COMPUTERIZED TREATMENT PLANNING;
D O I
10.1016/0167-8140(91)90085-U
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
In high energy X-ray beams, the dose calculated near the exit surface under electronic equilibrium conditions is generally overestimated since it is derived from measurements performed in water with large thickness of backscattering material. The resulting error depends on a number of parameters such as beam energy, field dimensions, thickness of overlying and underlying material. We have systematically measured for four different energies and for different combinations of the above parameters, the reduction of dose due to the lack of backscatter. This correction is expressed as a multiplicative factor, called "Backscatter Correction Factor" (BCF). This BCF is larger for lower energies, larger field sizes and larger depths. The BCF has been represented by an analytical expression which involves an exponential function of the backscattering thickness and linear relationships with depth, field size and beam quality index. Using this expression, the BCF can be calculated within 0.5% for any conditions in the energy range investigated.
引用
收藏
页码:128 / 134
页数:7
相关论文
共 14 条
[1]  
[Anonymous], 1983, MED PHYS, V10, P741
[2]   ON THE USE OF A QUALITY INDEX TO SPECIFY HIGH-ENERGY PHOTON BEAMS [J].
BRIDIER, A ;
BEAUVAIS, H ;
DUTREIX, A .
RADIOTHERAPY AND ONCOLOGY, 1986, 5 (01) :65-73
[3]   SURFACE AND EMERGENT DOSES IN RADIOTHERAPY [J].
CLARKSON, JR ;
HERBERT, RJT .
BRITISH JOURNAL OF RADIOLOGY, 1948, 21 (250) :494-500
[4]  
Gagnon W F, 1979, Med Phys, V6, P285, DOI 10.1118/1.594583
[5]  
GOEDE M R, 1977, Medical Physics (Woodbury), V4, P123, DOI 10.1118/1.594308
[6]  
JOHNS EH, 1974, PHYSICS RADIOLOGY, P374
[7]  
LAMPERT GD, 1983, BRIT J RADIOL, V56, P329
[9]   LETUDE DE LA DOSE AU VOISINAGE DES INTERFACES AIR-TISSU ET TISSU-AIR EN TELECOBALT-THERAPIE [J].
LEGARE, JM .
ACTA RADIOLOGICA-THERAPY PHYSICS BIOLOGY, 1964, 2 (02) :150-&
[10]   EXIT-SURFACE-DOSE CORRECTION FACTORS FOR X-RAYS OF 1.5 TO 4.0 MM CU HALF-VALUE LAYER [J].
LEGARE, JM .
RADIOLOGY, 1964, 82 (02) :272-279