AN IMPROVED NASAL PRONG APPARATUS FOR END-TIDAL CARBON-DIOXIDE MONITORING IN AWAKE, SEDATED PATIENTS

被引:21
作者
ROY, J
MCNULTY, SE
TORJMAN, MC
机构
[1] Department of Anesthesiology, Jefferson Medical College, Thomas Jefferson University, Philadelphia, 19107, PA
来源
JOURNAL OF CLINICAL MONITORING | 1991年 / 7卷 / 03期
关键词
MEASUREMENT TECHNIQUES; CAPNOGRAPHY; EQUIPMENT; CANNULAS; MONITORING; END-TIDAL;
D O I
10.1007/BF01619269
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
We describe and evaluate a new apparatus that monitors end-tidal carbon dioxide (PETCO2) and augments the inspired oxygen concentration in awake, sedated patients. The unit was evaluated for its effectiveness as an oxygenation device and its accuracy as a predictor of PaCO2 through the correlation of PaCO2 with PETCO2. Twenty cardiac surgical patients, physical status ASA 2-4, participated in this study. The PETCO2 monitoring device consisted of a dual-prong nasal oxygen cannula and a 14-gauge intravenous catheter that was inserted into one limb of the oxygen supply tubing and connected to a Datex gas analyzer (Datex Instrumentation Corp, Helsinki, Finland) to measure PETCO2. The cross-over passage between the prongs was intentionally blocked with the end of a wooden-core cotton swab. The oxygen flow rates were randomly varied (2, 4, and 6 L/min) every 5 minutes, and values for PETCO2 as well as arterial blood samples for analysis of PaCO2 and PaO2 were obtained at the end of each 5-minute period. The accuracy of the system was assessed by comparing the PaCO2-PETCO2 differences (bias) at each oxygen flow rate. The ratios of PETCO2 compared with PaCO2 were 0.98, 0.94, and 0.85, with correlation coefficients of r = 0.81, 0.85, and 0.63, respectively. The PaO2 values were 114, 154, and 183 mm Hg for the corresponding nasal oxygen flow rates of 2, 4, and 6 L/min, respectively. This study indicates that this modified nasal cannula provides supplemental oxygen adequately and yields a satisfactory reflection of the PaCO2 depending on the oxygen flow rate delivered.
引用
收藏
页码:249 / 252
页数:4
相关论文
共 13 条
[1]  
Ibarra E., Lees D.E., Mass spectrophotometer monitoring of patients with regional anesthesia, Anesthesiology, 63, pp. 572-573, (1985)
[2]  
Norman E.A., Zeig N.J., Ahmad I., Better designs for mass spectrometer monitoring of the awake patient, Anesthesiology, 64, (1986)
[3]  
Huntington C.T., King H.K., A simpler design for mass spectrometer monitoring of the awake patient, Anesthesiology, 65, pp. 565-566, (1986)
[4]  
Goldman J.M., A simple, easy, and inexpensive method for monitoring ETCO<sub>2</sub> through nasal cannulae, Anesthesiology, 67, (1987)
[5]  
Pressman M.A., A simple method of measuring ET-CO<sub>2</sub> during MAC and major regional anesthesia, Anesth Analg, 67, pp. 905-906, (1988)
[6]  
Louwsma D.L., Silverman D.G., Reproducibility of end tidal CO<sub>2</sub> measurements in sedated patients receiving supplemental O<sub>2</sub> by nasal cannula, Anesthesiology, 69, (1988)
[7]  
Bowe E.A., Boysen P.G., Broome J.A., Klein E.F., Accurate determination of end-tidal carbon dioxide during administration of oxygen by nasal cannulae, J Clin Monit, 5, pp. 105-110, (1989)
[8]  
Dunphy J.A., Accuracy of expired carbon dioxide partial pressure sampled from a nasal cannula II, Anesthesiology, 68, pp. 960-961, (1988)
[9]  
Barr P.O., Pulmonary gas exchange in man as affected by prolonged gravitational stress, Acta Physiol Scand Suppl, 207, pp. 1-46, (1963)
[10]  
Jones N.L., Robertson D.G., Kane J.W., Difference between end-tidal and arterial pCO<sub>2</sub>, J Appl Physiol, 47, pp. 954-960, (1979)