POTENTIAL HAZARDS OF RADIATIVE ELECTROMAGNETIC HYPERTHERMIA IN THE PRESENCE OF MULTIPLE METALLIC SURGICAL CLIPS

被引:11
作者
LEE, ER
SULLIVAN, DM
KAPP, DS
机构
[1] Department of Radiation Oncology, Stanford University School of Medicine, Stanford
关键词
SAR; METALLIC OBJECTS; 3-D MODELING; DEEP HYPERTHERMIA;
D O I
10.3109/02656739209005028
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Three-dimensional computer simulations of a patient under consideration for heating by the BSD Sigma 60 phased array applicator showed marked specific absorption rate (SAR) enhancement in the region where there was a concentration of metallic surgical clips. Subsequent tests in tissue-equivalent phantom material confirmed that metallic objects much shorter than the wavelength in tissue of the radiative electromagnetic field can cause significant changes in the power deposition near the implanted metal. The placement of extensive number of surgical clips may therefore hinder the ability to subsequently utilize radiative electromagnetic heating modalities.
引用
收藏
页码:809 / 817
页数:9
相关论文
共 13 条
[1]   HYPERTHERMIA IN CANCER-THERAPY - CURRENT STATUS .12. [J].
ANDERSON, RL ;
KAPP, DS .
MEDICAL JOURNAL OF AUSTRALIA, 1990, 152 (06) :310-315
[2]   PERIPHERAL NEUROPATHY AND MYONECROSIS FOLLOWING HYPERTHERMIA AND RADIATION-THERAPY FOR RECURRENT PROSTATIC-CANCER - CORRELATION OF DAMAGE WITH PREDICTED SAR PATTERN [J].
BENYOSEF, R ;
SULLIVAN, DM ;
KAPP, DS .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1992, 8 (02) :173-185
[4]   IFOSFAMIDE PLUS ETOPOSIDE COMBINED WITH REGIONAL HYPERTHERMIA IN PATIENTS WITH LOCALLY ADVANCED SARCOMAS - A PHASE-II STUDY [J].
ISSELS, RD ;
PRENNINGER, SW ;
NAGELE, A ;
BOEHM, E ;
SAUER, H ;
JAUCH, KW ;
DENECKE, H ;
BERGER, H ;
PETER, K ;
WILMANNS, W .
JOURNAL OF CLINICAL ONCOLOGY, 1990, 8 (11) :1818-1829
[5]   CREATION OF 3-DIMENSIONAL PATIENT MODELS FOR HYPERTHERMIA TREATMENT PLANNING [J].
JAMES, BJ ;
SULLIVAN, DM .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1992, 39 (03) :238-242
[6]  
JAMES BJ, 1992, IN PRESS IEEE T BIOM
[7]  
KAPP D S, 1986, International Journal of Hyperthermia, V2, P139, DOI 10.3109/02656738609012390
[8]   STANFORD-UNIVERSITY INSTITUTIONAL REPORT .1. EVALUATION OF EQUIPMENT FOR HYPERTHERMIA TREATMENT OF CANCER [J].
KAPP, DS ;
FESSENDEN, P ;
SAMULSKI, TV ;
BAGSHAW, MA ;
COX, RS ;
LEE, ER ;
LOHRBACH, AW ;
MEYER, JL ;
PRIONAS, SD .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1988, 4 (01) :75-115
[9]  
KAPP DS, 1991, IN PRESS INT J RAD O
[10]   A FREQUENCY-DEPENDENT FDTD METHOD FOR BIOLOGICAL APPLICATIONS [J].
SULLIVAN, DM .
IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 1992, 40 (03) :532-539