VOLTAMMETRIC BEHAVIOR OF ASCORBIC-ACID AT A GRAPHITE EPOXY COMPOSITE ELECTRODE CHEMICALLY MODIFIED WITH COBALT PHTHALOCYANINE AND ITS AMPEROMETRIC DETERMINATION IN MULTIVITAMIN PREPARATIONS

被引:112
作者
WRING, SA
HART, JP
BIRCH, BJ
机构
[1] BRISTOL POLYTECH,DEPT SCI,COLDHARBOUR LANE,BRISTOL BS16 1QY,ENGLAND
[2] UNILEVER RES LABS,COLWORTH LAB,SENSORS GRP,SHARNBROOK MK44 1LQ,BEDS,ENGLAND
关键词
D O I
10.1016/S0003-2670(00)85110-7
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Cyclic voltammetry was used to investigate the electrochemical behaviour of ascorbic acid at a carbon-epoxy composite electrode modified with the electron mediator cobalt phthalocyanine. The modified electrode reduced the overpotential necessary for the oxidation of the vitamin by approximately 150 mV to 0.21 V vs. The saturated calomel electrode; the process was dependent on the pH of the supporting electrolyte, but independent of ionic strength over the range studied. The relative standard deviation (r.s.d.) of the peak heights of the cyclic voltammograms was 0.81% for a 1 × 10-4 M ascorbic acid solution (n = 7). The optimum supporting electrolyte was found to be 0.05 M phosphate buffer (pH 5). Amperometry in stirred solutions was then done at an applied potential of +0.25 V. The limit of detection was 0.65 ng ml-1 and the calibration graph was linear in the range 175 ng ml-1-50 μg ml-1. The method was used to determine ascorbic acid concentrations in single- and multivitamin preparations; the recovery was 97.86% for the vitamin added to one preparation. The r.s.d. for the analyses in these samples was about 5%. For comparison, the vitamin was also determined in these tablets using LC with UV detection at 254 nm; the correlation coefficient for the levels determined was 0.9989 (p = 0.0007). © 1990.
引用
收藏
页码:63 / 70
页数:8
相关论文
共 21 条
[1]  
BARD AJ, 1980, ELECTROCHEMICAL METH, P123
[2]  
BOND AM, 1980, MODERN POLAROGRAPHIC, P194
[3]   VACUUM HEAT-TREATMENT FOR ACTIVATION OF GLASSY-CARBON ELECTRODES [J].
FAGAN, DT ;
HU, IF ;
KUWANA, T .
ANALYTICAL CHEMISTRY, 1985, 57 (14) :2759-2763
[4]   RESOLUTION OF ASCORBIC, DEHYDROASCORBIC AND DIKETOGULONIC ACIDS BY PAIRED-ION REVERSED-PHASE CHROMATOGRAPHY [J].
FINLEY, JW ;
DUANG, E .
JOURNAL OF CHROMATOGRAPHY, 1981, 207 (03) :449-453
[5]  
FOOT EA, 1982, ANN CLIN BIOCH S1, V24, pS1
[6]   CHEMICALLY MODIFIED CARBON PASTE ELECTRODES .4. ELECTROSTATIC BINDING AND ELECTROCATALYSIS AT POLY(4-VINYLPYRIDINE)-CONTAINING ELECTRODES [J].
GENO, PW ;
RAVICHANDRAN, K ;
BALDWIN, RP .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1985, 183 (1-2) :155-166
[7]   ELECTROSORPTION OF VITAMIN-K1 AT MERCURY AND ITS DETERMINATION AT SUB-MICROGRAM LEVELS BY DIFFERENTIAL PULSE VOLTAMMETRY AT A HANGING MERCURY-ELECTRODE [J].
HART, JP ;
CATTERALL, A .
ANALYTICA CHIMICA ACTA, 1981, 128 (JUL) :245-250
[8]  
HART JP, 1984, INVESTIGATIVE MICROT, V1, P199
[9]  
KUTIK MA, 1982, ANAL BIOCHEM, V66, P424
[10]   SIMULTANEOUS DETERMINATION OF ASCORBIC-ACID, ISOASCORBIC ACID (ERYTHORBIC ACID) AND URIC-ACID IN HUMAN-PLASMA BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY WITH AMPEROMETRIC DETECTION [J].
KUTNINK, MA ;
SKALA, JH ;
SAUBERLICH, HE ;
OMAYE, ST .
JOURNAL OF LIQUID CHROMATOGRAPHY, 1985, 8 (01) :31-46