RANGE VELOCITY LIMITATIONS FOR TIME-DOMAIN BLOOD VELOCITY ESTIMATION

被引:29
作者
JENSEN, JA
机构
[1] Bioacoustics Research Laboratory, Department of Electrical and Computer Engineering, University of Illinois at Urbana-Champaign, Urbana, IL 61801-2991
关键词
VELOCITY ESTIMATION; MEDICAL ULTRASOUND; RANGE VELOCITY LIMITATION;
D O I
10.1016/0301-5629(93)90091-2
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
The traditional range/velocity limitation for blood velocity estimation systems using ultrasound is elucidated. It is stated that the equation is a property of the estimator used, not the actual physical measurement situation, as higher velocities can be estimated by the time domain cross-correlation approach. It is demonstrated that the time domain technique under certain measurement conditions will yield unsatisfactory results, when trying to estimate high velocities. Various methods to avoid these artifacts using temporal and spatial clustering techniques are suggested. The improvement in probability of correct detection is derived, and several examples of simulations are shown.
引用
收藏
页码:741 / 749
页数:9
相关论文
共 12 条
[1]  
Bonnefous O., 1986, IEEE 1986 Ultrasonics Symposium Proceedings (Cat. No.86CH2375-4), P855
[2]  
BONNEFOUS O, 1989, IEEE 1989 ULTRASONICS SYMPOSIUM : PROCEEDINGS, VOLS 1 AND 2, P887, DOI 10.1109/ULTSYM.1989.67116
[3]  
DOTTI D, 1976, ENERG NUCL-MILAN, V23, P571
[4]  
Evans D. H., 1989, DOPPLER ULTRASOUND P
[6]   FLOW VELOCITY PROFILE VIA TIME-DOMAIN CORRELATION ERROR ANALYSIS AND COMPUTER-SIMULATION [J].
FOSTER, SG ;
EMBREE, PM ;
OBRIEN, WD .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1990, 37 (03) :164-175
[7]  
Hatle L, 1982, DOPPLER ULTRASOUND C
[8]  
JENSEN JA, 1993, IN PRESS MED BIO ENG
[9]  
JENSEN JA, 1993, IN PRESS IEEE T MED
[10]   REAL-TIME TWO-DIMENSIONAL BLOOD-FLOW IMAGING USING AN AUTO-CORRELATION TECHNIQUE [J].
KASAI, C ;
NAMEKAWA, K ;
KOYANO, A ;
OMOTO, R .
IEEE TRANSACTIONS ON SONICS AND ULTRASONICS, 1985, 32 (03) :458-464