COMPARISON OF 2 AUTOMATED NEPHELOMETERS

被引:7
作者
EMANCIPATOR, K [1 ]
ELIN, RJ [1 ]
FLEISHER, TA [1 ]
机构
[1] NIH,WARREN G MAGNUSON CLIN CTR,DEPT CLIN PATHOL,BETHESDA,MD 20892
关键词
IMMUNOGLOBULIN; APOLIPOPROTEIN; COMPLEMENT; ACUTE PHASE REACTANTS; ANTITRYPSIN; CERULOPLASMIN; ALBUMIN; C-REACTIVE PROTEIN; STANDARDIZATION; PROTEIN ASSAY;
D O I
10.1002/jcla.1860060611
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
The intercalibration precision and linearity were determined for two representative analytes, apolipoprotein A1 (apo A1) and immunoglobulin G (IgG), on the Beckman Array(R) and the Behring Nephelometer 100(R) (BN-100). For two of nine samples analyzed, poor precision was observed for IgG with the Array. The poor precision for these two samples is attributed to large systematic shifts. A statistical non-linearity was observed for apo A1 with the BN-100, but this non-linearity does not exclude use of this assay for clinical diagnosis. Method comparisons were done for 12 analytes: apo A1, apo B, IgG, IgA, IgM, IgE, C3, C4, albumin, C-reactive protein, alpha-1-antitrypsin, and ceruloplasmin. These comparisons showed proportional biases of > 10% for seven of 12 analytes. Furthermore, correlation coefficients were <0.96 for seven of 12 analytes. We conclude that comparing results obtained from the two different nephelometers is of only limited value for the individual patient.
引用
收藏
页码:399 / 404
页数:6
相关论文
共 7 条
[1]  
CORNBLEET PJ, 1979, CLIN CHEM, V25, P432
[2]  
DAVID G, 1990, EP EVALUATOR RELEASE
[3]  
MANDEL J, 1964, STATISTICAL ANAL EXP, P288
[4]   DETERMINATION OF 5 PROTEIN ANALYTES USING THE BECKMAN ARRAY AND THE BEHRING NEPHELOMETER SYSTEM [J].
SCHOTTERS, SB ;
MCBRIDE, JH ;
RODGERSON, DO ;
HIGGINS, S ;
PISA, M .
JOURNAL OF CLINICAL LABORATORY ANALYSIS, 1988, 2 (02) :108-111
[5]  
1988, BECKMAN ARRAY PROTEI, P1
[6]  
1989, BEHRING NEPHELOMETER, P1
[7]  
1986, EVALUATION LINEARITY