CORONARY-ARTERY IMAGING AT 0.5 T USING SEGMENTED 3D ECHO-PLANAR IMAGING

被引:50
作者
BORNERT, P
JENSEN, D
机构
[1] Philips Research, Department Technical Systems, Hamburg
关键词
MAGNETIC RESONANCE IMAGING; CORONARY ARTERY IMAGING; EPI; 3D;
D O I
10.1002/mrm.1910340602
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The application of segmented 3D gradient echo EPI at 0.5 T for coronary artery imaging is described. Experiments were performed using fat suppression, ECG triggering, and a patient-controlled breath-holding scheme. This approach provides a sufficient signal-to-noise ratio for thin contiguous slices in conjunction with a phased array cardiac receive coil. Wide 3D volumes, covering the proximal branches of the coronary tree, were measured with a high spatial resolution. Such data sets can be used for subsequent vessel segmentation. Furthermore, data out of narrow 3D volumes were obtained containing fewer slices angulated in the direction of a selected coronary artery, This provides a good visualization of the selected vessel over several centimeters without the need for segmentation.
引用
收藏
页码:779 / 785
页数:7
相关论文
共 23 条
  • [1] BORNERT P, 1994, 2ND P SMR ANN M SAN, P372
  • [2] BORNERT P, 1994, P SMR WORKSHOP CARDI, P3
  • [3] CORONARY ANGIOGRAPHY WITH MAGNETIZATION-PREPARED T-2 CONTRAST
    BRITTAIN, JH
    HU, BS
    WRIGHT, GA
    MEYER, CH
    MACOVSKI, A
    NISHIMURA, DG
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1995, 33 (05) : 689 - 696
  • [4] CARLSEN IC, 1994, 2ND P SMR ANN M SAN, P834
  • [5] REAL-TIME MOVIE IMAGING FROM A SINGLE CARDIAC CYCLE BY NMR
    CHAPMAN, B
    TURNER, R
    ORDIDGE, RJ
    DOYLE, M
    CAWLEY, M
    COXON, R
    GLOVER, P
    MANSFIELD, P
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1987, 5 (03) : 246 - 254
  • [6] Cho ZH, 1987, 6TH P SMRM ANN M, P912
  • [7] CORONARY-ARTERIES - BREATH-HOLD MR ANGIOGRAPHY
    EDELMAN, RR
    MANNING, WJ
    BURSTEIN, D
    PAULIN, S
    [J]. RADIOLOGY, 1991, 181 (03) : 641 - 643
  • [8] HALVING MR IMAGING TIME BY CONJUGATION - DEMONSTRATION AT 3.5 KG
    FEINBERG, DA
    HALE, JD
    WATTS, JC
    KAUFMAN, L
    MARK, A
    [J]. RADIOLOGY, 1986, 161 (02) : 527 - 531
  • [9] ECHO-PLANAR HIGH-RESOLUTION FLOW VELOCITY MAPPING
    FIRMIN, DN
    KLIPSTEIN, RH
    HOUNSFIELD, GL
    PALEY, MP
    LONGMORE, DB
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1989, 12 (03) : 316 - 327