DETERMINATION OF CATECHOLAMINES IN URINE BY LIQUID-CHROMATOGRAPHY AND ELECTROCHEMICAL DETECTION AFTER ONLINE SAMPLE PURIFICATION ON IMMOBILIZED BORONIC ACID

被引:21
作者
ERIKSSON, BM
WIKSTROM, M
机构
来源
JOURNAL OF CHROMATOGRAPHY | 1992年 / 593卷 / 1-2期
关键词
D O I
10.1016/0021-9673(92)80285-3
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Norepinephrine, epinephrine and dopamine in urine were measured by an automated liquid chromatographic method. After sample purification on a column containing silica-immobilized boronic acid, which showed great affinity for catecholamines at neutral pH, the catecholamines were eluted by backflushing with an acidic mobile phase and transferred to a cation exchanger for separation. Detection was performed electrochemically and the relative standard deviation was 2% for the analysis of endogenous concentrations in human urine.
引用
收藏
页码:185 / 190
页数:6
相关论文
共 21 条