ANTICOAGULANT POISONING IN ANIMALS - A SIMPLE NEW HIGH-PERFORMANCE THIN-LAYER CHROMATOGRAPHIC (HPTLC) METHOD FOR THE SIMULTANEOUS DETERMINATION OF 8 ANTICOAGULANT RODENTICIDES IN LIVER SAMPLES

被引:43
作者
BERNY, PJ
BURONFOSSE, T
LORGUE, G
机构
[1] Département des Sciences Pharmaceutiques et Toxicologique, Ecole Nationale Vététrinaire de Lyon, Marcy l’Etoile, F-69280
关键词
D O I
10.1093/jat/19.7.576
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The purpose of this study was to develop and evaluate a technique for the analysis of anticoagulant rodenticides in serum and liver samples using a new high-performance thin-layer chromatographic apparatus. Detection limits were estimated at 0.2 µg/g in liver extracts for eight different substances. Overall, this technique was repeatable and reproducible. The percent recovery was greater than 87% for each substance. Liver and serum samples of animals known to be exposed to one anticoagulant and showing clinical signs of poisoning were analyzed. Concentrations measured varied between 0.2 and 3 µg/g (liver extracts). Only blood samples from one dog could be analyzed. The concentration was 150 ng/mL the first day after admission and 140 ng/mL the following day. Analyses are technically easily and rapidly performed, and they are inexpensive. Therefore, this technique could be a valuable alternative to current high-performance liquid chromatographic methods. © 1995, Oxford University Press.
引用
收藏
页码:576 / 580
页数:5