FLOW-INJECTION ANALYSIS OF PHENOLS AT A GRAPHITE ELECTRODE MODIFIED WITH CO-IMMOBILIZED LACCASE AND TYROSINASE

被引:146
作者
YAROPOLOV, AI [1 ]
KHARYBIN, AN [1 ]
EMNEUS, J [1 ]
MARKOVARGA, G [1 ]
GORTON, L [1 ]
机构
[1] LUND UNIV,DEPT ANALYT CHEM,S-22100 LUND,SWEDEN
关键词
AMPEROMETRY; BIOSENSORS; FLOW INJECTION; LACCASE; TYROSINASE; PHENOLS;
D O I
10.1016/0003-2670(94)00404-A
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The kinetic parameters of the oxidation reaction of phenolic compounds by molecular oxygen catalysed by fungal laccase have been studied. An amperometric biosensor for detection of phenols in environmental analysis is proposed. The enzymes laccase and tyrosinase were co-immobilised by adsorption onto a spectrographic graphite electrode. The bienzyme electrode was used as a sensor in a single-line flow-injection system. The analysis is based on the amperometric detection of the enzymatic reaction products at a potential of -0.05 V vs. Ag/AgCl. The experimental parameters were optimised. The joint use of laccase and tyrosinase in the analytical procedure allows the detection of a large group of phenolic compounds.
引用
收藏
页码:137 / 144
页数:8
相关论文
共 34 条
[1]   BIOAMPEROMETRIC SENSORS FOR PHENOL BASED ON CARBON PASTE ELECTRODES [J].
BONAKDAR, M ;
VILCHEZ, JL ;
MOTTOLA, HA .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1989, 266 (01) :47-55
[2]   ULTRAVIOLET SPECTROPHOTOMETRIC DETERMINATION OF PHENOLS IN NATURAL AND WASTE-WATERS WITH IODINE MONOBROMIDE [J].
BOSCH, F ;
FONT, G ;
MANES, J .
ANALYST, 1987, 112 (09) :1335-1337
[3]   DETERMINATION OF PHENOL IN WASTES AND WATER USING AN ENZYME SENSOR [J].
CAMPANELLA, L ;
BEONE, T ;
SAMMARTINO, MP ;
TOMASSETTI, M .
ANALYST, 1993, 118 (08) :979-986
[4]   SILICONE-GREASE-BASED IMMOBILIZATION METHOD FOR THE PREPARATION OF ENZYME ELECTRODES [J].
CONNOR, MP ;
SANCHEZ, J ;
WANG, J ;
SMYTH, MR ;
MANNINO, S .
ANALYST, 1989, 114 (11) :1427-1429
[5]  
DOMINGUEZ E, 1994, COMMUNICATION
[6]  
DUCKWORTH HW, 1970, J BIOL CHEM, V245, P1613
[7]   LACCASE-BASED BIOSENSOR FOR DETERMINATION OF POLYPHENOLS - DETERMINATION OF CATECHOLS IN TEA [J].
GHINDILIS, AL ;
GAVRILOVA, VP ;
YAROPOLOV, AI .
BIOSENSORS & BIOELECTRONICS, 1992, 7 (02) :127-131
[8]  
GHINDILIS AL, 1988, BIOCHEMISTRY-US, V53, P635
[9]  
GORTON L, 1994, IN PRESS ELECTROANAL
[10]   THE DETERMINATION OF P-CRESOL IN CHLOROFORM WITH AN ENZYME ELECTRODE USED IN THE ORGANIC-PHASE [J].
HALL, GF ;
BEST, DJ ;
TURNER, APF .
ANALYTICA CHIMICA ACTA, 1988, 213 (1-2) :113-119