APPLICATION OF A POLY(DITHIOCARBAMATE) RESIN WITH MACRORETICULAR SUPPORT TO THE DETERMINATION OF TRACE AMOUNTS OF CADMIUM AND LEAD IN NONSALINE WATERS

被引:24
作者
YEBRABIURRUN, MC
BERMEJOBARRERA, A
BERMEJOBARRERA, MP
机构
关键词
PRECONCENTRATION; POLY(DITHIOCARBAMATE) RESIN; CADMIUM AND LEAD; NONSALINE WATER; FLAME ATOMIC ABSORPTION SPECTROMETRY;
D O I
10.1039/an9911601033
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The chelating poly(dithiocarbamate) resin with macroreticular support is shown to be effective for the preconcentration of cadmium and lead from non-saline waters. The retained ions can be eluted efficiently in 5 ml of 8 mol dm-3 nitric acid from 1-20 cm columns of resin. Poly(dithiocarbamate) resin is used for preconcentration in the determination of cadmium and lead by flame atomic absorption spectrometry. The common ions present in non-saline waters do not interfere. The sensitivity was 0.10-mu-g l-1 of Cd and 0.73-mu-g l-1 of Pb. The detection limits were 0.03 and 0.34-mu-g l-1 for Cd and Pb, respectively. The cadmium and lead concentration ranges, determined in mineral and tap waters, were between 0.1 and 0.8-mu-g l-1 and between 0.7 and 16.9-mu-g l-1, respectively.
引用
收藏
页码:1033 / 1035
页数:3