A GENERALIZED GIBBS PRIOR FOR MAXIMUM A-POSTERIORI RECONSTRUCTION IN SPECT

被引:39
作者
LALUSH, DS [1 ]
TSUI, BMW [1 ]
机构
[1] UNIV N CAROLINA, DEPT RADIOL, CHAPEL HILL, NC 27599 USA
关键词
D O I
10.1088/0031-9155/38/6/007
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In this paper, we introduce a Gibbs prior for use in MAP (maximum a posteriori) reconstruction in SPECT. This new prior, with three parameters, is able to approximate the results of previously-proposed priors with two parameters, as well as a continuum of others. Also, it allows the user increased flexibility in selecting the properties to be emphasized in the final reconstructed image estimate. The additional flexibility offered by the new prior is important in addressing the problem of selecting a prior and its associated parameters in a clinical situation. The paper demonstrates the importance of the derivative potential function (DPF) of the Gibbs distribution in determining which properties will be emphasized in the iterated image estimates. The effects of each of the three parameters are demonstrated on reconstructions from acquired SPECT data. We conclude that the parameters must be chosen carefully with consideration for the object distribution and the relative requirements for low-contrast detail, smoothing and edge sharpness in the reconstructed image.
引用
收藏
页码:729 / 741
页数:13
相关论文
共 14 条
[1]   STOCHASTIC RELAXATION, GIBBS DISTRIBUTIONS, AND THE BAYESIAN RESTORATION OF IMAGES [J].
GEMAN, S ;
GEMAN, D .
IEEE TRANSACTIONS ON PATTERN ANALYSIS AND MACHINE INTELLIGENCE, 1984, 6 (06) :721-741
[2]  
Geman S., 1985, P STAT COMP SECT AM, V21, P12
[3]  
GILLIAD DR, 1988, P NATO ADV STUDY I F, P525
[4]  
GRADSHTEYN IS, 1980, TABLE INTEGRALS SERI, P80
[5]   BAYESIAN RECONSTRUCTIONS FROM EMISSION TOMOGRAPHY DATA USING A MODIFIED EM ALGORITHM [J].
GREEN, PJ .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1990, 9 (01) :84-93
[6]   A GENERALIZED EM ALGORITHM FOR 3-D BAYESIAN RECONSTRUCTION FROM POISSON DATA USING GIBBS PRIORS [J].
HEBERT, T ;
LEAHY, R .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1989, 8 (02) :194-202
[7]   3-D PHANTOM TO SIMULATE CEREBRAL BLOOD-FLOW AND METABOLIC IMAGES FOR PET [J].
HOFFMAN, EJ ;
CUTLER, PD ;
DIGBY, WM ;
MAZZIOTTA, JC .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1990, 37 (02) :616-620
[8]  
JOHNSON VE, 1991, PROG CLIN BIOL RES, V363, P15
[9]   SIMULATION EVALUATION OF GIBBS PRIOR DISTRIBUTIONS FOR USE IN MAXIMUM A-POSTERIORI SPECT RECONSTRUCTIONS [J].
LALUSH, DS ;
TSUI, BMW .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1992, 11 (02) :267-275
[10]  
LALUSH DS, 1992, THESIS U N CAROLINA