ACCURACY OF DIFFERENT ROUTES OF INTRACRANIAL-PRESSURE (ICP) MONITORING USING MICROTRANSDUCERS
被引:1
作者
:
POON, WS
论文数:
0
引用数:
0
h-index:
0
机构:
CHINESE UNIV HONG KONG,PRINCE WALES HOSP,DEPT SURG,SHA TIN,HONG KONG
CHINESE UNIV HONG KONG,PRINCE WALES HOSP,DEPT SURG,SHA TIN,HONG KONG
POON, WS
[
1
]
LI, AKC
论文数:
0
引用数:
0
h-index:
0
机构:
CHINESE UNIV HONG KONG,PRINCE WALES HOSP,DEPT SURG,SHA TIN,HONG KONG
CHINESE UNIV HONG KONG,PRINCE WALES HOSP,DEPT SURG,SHA TIN,HONG KONG
LI, AKC
[
1
]
机构
:
[1]
CHINESE UNIV HONG KONG,PRINCE WALES HOSP,DEPT SURG,SHA TIN,HONG KONG
来源
:
JOURNAL OF NEUROSURGICAL ANESTHESIOLOGY
|
1990年
/ 2卷
/ 04期
关键词
:
D O I
:
10.1097/00008506-199012000-00005
中图分类号
:
R614 [麻醉学];
学科分类号
:
100217 ;
摘要
:
Intracranial pressure (ICP) monitoring using microtransducer and intraventricular pressure (IVP) was compared simultaneously in 20 consecutive head-injured patients within a 1-year period. In the first four patients with subdurally placed transducers, an acceptable corresponding rate with IVP was demonstrated (88%). In the remaining 16 patients with extradurally placed transducers, acceptable accuracy (73%) was only achieved in the last four. This accuracy was repeated in an additional six patients with extradurally placed transducers (75% corresponding with IVP). We have therefore suggested that, before extradural pressure monitoring can be employed alone in clinical practice, a comparison with IVP should be performed. © Lippincott-Raven Publishers.
引用
收藏
页码:280 / 283
页数:4
相关论文
未找到相关数据
未找到相关数据