ACCURACY OF DIFFERENT ROUTES OF INTRACRANIAL-PRESSURE (ICP) MONITORING USING MICROTRANSDUCERS

被引:1
作者
POON, WS [1 ]
LI, AKC [1 ]
机构
[1] CHINESE UNIV HONG KONG,PRINCE WALES HOSP,DEPT SURG,SHA TIN,HONG KONG
关键词
D O I
10.1097/00008506-199012000-00005
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Intracranial pressure (ICP) monitoring using microtransducer and intraventricular pressure (IVP) was compared simultaneously in 20 consecutive head-injured patients within a 1-year period. In the first four patients with subdurally placed transducers, an acceptable corresponding rate with IVP was demonstrated (88%). In the remaining 16 patients with extradurally placed transducers, acceptable accuracy (73%) was only achieved in the last four. This accuracy was repeated in an additional six patients with extradurally placed transducers (75% corresponding with IVP). We have therefore suggested that, before extradural pressure monitoring can be employed alone in clinical practice, a comparison with IVP should be performed. © Lippincott-Raven Publishers.