KETONE INTERFERENCE IN ESTIMATION OF URINARY CREATININE - EFFECT ON CREATININE CLEARANCE IN DIABETIC KETOCIDOSIS

被引:11
作者
BLANK, DW
NANJI, AA
机构
[1] VANCOUVER GEN HOSP, DIV CLIN CHEM, VANCOUVER V5Z 1M9, BC, CANADA
[2] UNIV BRITISH COLUMBIA, DEPT PATHOL, VANCOUVER V6T 1W5, BC, CANADA
关键词
D O I
10.1016/S0009-9120(82)96774-1
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Acetoacetate is known to spuriously raise serum creatinine concentration in patients with diabetic ketoacidosis. Its effect on urinary creatinine has not been studied. Since the renal threshold for ketoacids is low, large amounts of acetoacetate may be present in the urine of uncontrolled diabetics. This interference was investigated using 3 different automated analyzers. It was found that +3 or +4 reactions with Ketostix, equivalent to greater than 10-15 mmol/l of acetoacetate caused significant interference with the Abbott VP and Beckman Astra instruments. This could cause errors in the calculation of creatinine clearance especially when serum creatinine is close to a normal level. It is recommended that measurement of creatinine clearance be delayed until better diabetic control is achieved or the creatinine be measured by a method which is free from ketone interference, e.g., by the Dupont aca.
引用
收藏
页码:279 / 280
页数:2
相关论文
共 5 条