EARLY RESULTS OF INTRAVASCULAR OXYGENATION

被引:5
作者
KALLIS, P
ALSAADY, NM
BENNETT, ED
TREASURE, T
机构
[1] Department of Cardiothoracic Surgery, St Georges Hospital
[2] Department of Intensive Care, St Georges Hospital
关键词
RESPIRATORY FAILURE; INTRAVASCULAR OXYGENATION;
D O I
10.1016/1010-7940(93)90160-D
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We have used the intravascular oxygenator (IVOX) in eight patients (14 - 76 years) with acute respiratory failure. At the time of referral for IVOX all patients were on inotropic support, five had pneumothoraces and two were on haemofiltration. Ventilatory data before IVOX were: ventilated for 2 - 14 days (median 6), PEEP 5 - 12 cm H2O (12), PIP 34 - 95 cm H2O (49), FiO2 0.7 - 1 (1), PaO2 5.4 - 26 KPa (8.5) and PaCO2 3.7 - 23 KPa (8). The intravascular oxygenator was used for 6 h - 12 days (median 4), the oxygen transfer achieved was 68 - 140 ml/min (85) and the carbon dioxide removed was 40 - 106 ml/min (59). It was possible to decrease the ventilator settings with a significant increase in the PaO2 by 48 h (P < 0.05) but the fall in PaCO2 did not reach significance. As the IVOX was turned from ''off'' to ''on'' the mean PaO2 increase was 0.35 KPa +/- 0.14 SEM (P = 0.04) and the mean PaCO2 decrease 0.7 KPa +/- 0.2 SEM (P = 0.02) without any significant change in cardiac output. Two patients survived and six died from multisystem failure with three patients being hypoxic at the time of death. Postmortem examinations on five patients did not reveal any IVOX-related complications. We conclude that IVOX is safe but it is not as efficient in gas exchange as extracorporeal membrane oxygenation at present. Further improvement in the gas exchange efficiency of this prototype could render IVOX a very useful device.
引用
收藏
页码:206 / 210
页数:5
相关论文
共 15 条
[1]  
Bartlett R.H., Rews A.F., Extracorporeal membrane oxygenation ECMO for newborn respiratory failure 45 cases, Surgery, 921, pp. 425-433, (1982)
[2]  
Dutton R.C., Mather F., Walker S.N., Lipps B.J., Rudy L.W., Severingaus J.W., Edmunds L.H., Development and evaluation of a new hollow fiber membrane oxygenator, Trans am Sot Artif Int Organs, 17, pp. 331-335, (1971)
[3]  
Gattinoni L., Pesenti A., Mascheroni D., Marcolin R., Fumagalli R., Rossi F., Iapichino G., Romagnoli G., Uziel L., Agostoni A., Kolobow T., Damia G., Low frequency positive pressure ventilation with extracorporeal CO removal in severe acute respiratory failure, JAMA, 256, pp. 881-886, (1986)
[4]  
Gibbon J.H., Application of a mechanical heart and lung apparatus to cardiac surgery, Minn Med, 37, pp. 171-180, (1954)
[5]  
Hill J.D., O'Brien T.G., Murray J.J., Dontigny L., Bramson M.L., Osborn J.J., Gerbode F., Extracorporeal oxygenation for acute post-traumatic respiratory failure shock lung syndrome use of the Bramson Membrane Lung, N Engl J Med, 286, pp. 1629-1634, (1972)
[6]  
Kallis P., Al-Saady N.M., Bennett E.D., Treasure T., Intravascular oxygenation with IVOX, Br J Hosp Med, 47, pp. 824-828, (1992)
[7]  
Mortensen J.D., An intravenacaval blood gas exchange IVCBGE device a preliminary report, Trans am Sot Artif Int Org, 33, pp. 570-573, (1987)
[8]  
Mortensen J.D., Berry G., Conceptual and design features of a practical, clinically effective, intravenous, mechanical blood oxygen/carbon dioxide exchange device IVOX, Int J Artif Organs, 12, pp. 384-389, (1989)
[9]  
Mortensen J.D., Talbot S., Burkart J.A., Cross-section internal diameters of human cervical and femoral blood vessels relationship to subjects sex, age, body size, Anat Ret, 225, pp. 115-124, (1990)
[10]  
Rashkind W.J., Freeman A., Klein D., Toft R.W., Evaluation of a disposable plastic, low volume, pumpless oxygenator as a lung substitute, J Pediatr, 66, pp. 94-102, (1965)