EFFECT OF MOTION ARTIFACT ON PULSE OXIMETERS - EVALUATION OF 4 INSTRUMENTS AND FINGER PROBES

被引:53
作者
LANGTON, JA
HANNING, CD
机构
[1] University Department of Anaesthesia, Leicester General Hospital
关键词
Equipment: pulse oximeters; Measurement techniques: pulse oximetry; motion artefact;
D O I
10.1093/bja/65.4.564
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
The ability of the Ohmeda 3700, Nellcor N200, Datex Satlite Plus and Simed S100 pulse oximeters to detect induced hypoxaemia in the presence of motion artefact was assessed, under conditions of controlled vibration using an industrial vibration facility. Vibration at 4 Hz and 8 Hz induced increases in detection time for hypoxaemia and spurious decreases in the displayed SaO2 in some of the oximeters tested. Finger-dependent differences in oximeter performance and pulse rate registration were noted especially in those oximeters without ECG linkage (Ohmeda 3700 and Simed S100). Subsequently, eight different pulse oximeter finger probes were assessed for those characteristics that may predispose to motion artefact. There were marked differences in the mass of the probes, the forces exerted on the test finger and in the force required to displace the probes from the subject's finger. Differences in both the microprocessor programmes and the physical characteristics of the finger probes may explain the observed differences in function. Similar studies should form part of the standard evaluation of new pulse oximeters. © 1990 Copyright: 1990 British Journal of Anaesthesia.
引用
收藏
页码:564 / 570
页数:7
相关论文
共 8 条
[1]  
ALEXANDER CM, 1989, ANESTH ANALG, V68, P368
[2]  
HUCH A, 1988, LANCET, V1, P357
[3]   PULSE OXIMETERS - ESSENTIAL MONITORS WITH LIMITATIONS [J].
KIDD, JF ;
VICKERS, MD .
BRITISH JOURNAL OF ANAESTHESIA, 1989, 62 (04) :355-357
[4]   COMPARISON OF 4 PULSE OXIMETERS - EFFECTS OF VENOUS OCCLUSION AND COLD-INDUCED PERIPHERAL VASOCONSTRICTION [J].
LANGTON, JA ;
LASSEY, D ;
HANNING, CD .
BRITISH JOURNAL OF ANAESTHESIA, 1990, 65 (02) :245-247
[5]   BLOOD-FLOW LIMITS AND PULSE OXIMETER SIGNAL-DETECTION [J].
LAWSON, D ;
NORLEY, I ;
KORBON, G ;
LOEB, R ;
ELLIS, J .
ANESTHESIOLOGY, 1987, 67 (04) :599-603
[6]   A COMPARISON OF 15 PULSE OXIMETERS .1. A CLINICAL COMPARISON .2. A TEST OF PERFORMANCE UNDER CONDITIONS OF POOR PERFUSION [J].
MORRIS, RW ;
NAIRN, M ;
TORDA, TA .
ANAESTHESIA AND INTENSIVE CARE, 1989, 17 (01) :62-73
[7]   HUMAN STUDIES CONCERNING THERMAL-INDUCED SHIVERING, POSTOPERATIVE SHIVERING, AND COLD-INDUCED VASODILATION [J].
POZOS, RS ;
ISRAEL, D ;
MCCUTCHEON, R ;
WITTMERS, LE ;
SESSLER, D .
ANNALS OF EMERGENCY MEDICINE, 1987, 16 (09) :1037-1041
[8]   COMPARISON OF PULSE OXIMETERS - EFFECTS OF VASOCONSTRICTION AND VENOUS ENGORGEMENT [J].
WILKINS, CJ ;
MOORES, M ;
HANNING, CD .
BRITISH JOURNAL OF ANAESTHESIA, 1989, 62 (04) :439-444