FAILURE OF A TITANIUM ANTERIOR CERVICAL PLATE IMPLANT - MICROSTRUCTURAL ANALYSIS OF FAILURE - CASE-REPORT

被引:6
作者
BALDWIN, NG
HARTMAN, GP
WEISER, MW
BENZEL, EC
机构
[1] UNIV NEW MEXICO,SCH MED,DEPT ORTHOPAED SURG,ALBUQUERQUE,NM 87131
[2] UNIV NEW MEXICO,DEPT MECH ENGN,ALBUQUERQUE,NM
关键词
INTERNAL FIXATION; SPINAL FUSION; SPINAL IMPLANTS; SPINAL INSTRUMENTATION;
D O I
10.3171/jns.1995.83.4.0741
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
The authors present the case of a failed titanium anterior cervical spine plate. The plate was contoured during implantation. Analysis of the failed implant with electron microscopy showed no metallurgical flaw to have caused the plate fracture. A crack was apparently initiated with the contouring of the plate. It was located at the weakest point in the plate (the position of the intermediate screw hole), and with subsequent stress, it propagated and led to fracture. Techniques for avoiding failure of this type are discussed.
引用
收藏
页码:741 / 743
页数:3
相关论文
共 10 条
[1]
BIOMECHANICAL EVALUATION OF CERVICAL SPINAL STABILIZATION METHODS IN A HUMAN CADAVERIC MODEL [J].
COE, JD ;
WARDEN, KE ;
SUTTERLIN, CE ;
MCAFEE, PC .
SPINE, 1989, 14 (10) :1122-1131
[2]
COMPLICATIONS OF CERVICAL-SPINE SURGERY - A 5-YEAR REPORT ON A SURVEY OF THE MEMBERSHIP OF THE CERVICAL-SPINE-RESEARCH-SOCIETY BY THE MORBIDITY-AND-MORTALITY-COMMITTEE [J].
GRAHAM, JJ .
SPINE, 1989, 14 (10) :1046-1050
[3]
FAILURE OF SYNTHES ANTERIOR CERVICAL FIXATION DEVICE BY FRACTURE OF MORSCHER SCREWS - A BIOMECHANICAL STUDY [J].
HOLLOWELL, JP ;
REINARTZ, J ;
PINTAR, FA ;
MORGESE, V ;
MAIMAN, DJ .
JOURNAL OF SPINAL DISORDERS, 1994, 7 (02) :120-125
[4]
PULL-OUT STRENGTH OF CASPAR CERVICAL SCREWS [J].
MAIMAN, DJ ;
PINTAR, FA ;
YOGANANDAN, N ;
REINARTZ, J ;
TOSELLI, R ;
WOODWARD, E ;
HAID, R .
NEUROSURGERY, 1992, 31 (06) :1097-1101
[5]
BIOMECHANICS OF CERVICAL-SPINE INTERNAL-FIXATION [J].
MONTESANO, PX ;
JUACH, EC ;
ANDERSON, PA ;
BENSON, DR ;
HANSON, PB .
SPINE, 1991, 16 (03) :S10-S16
[6]
ESOPHAGEAL-PERFORATION FOLLOWING ANTERIOR CERVICAL-SPINE SURGERY [J].
NEWHOUSE, KE ;
LINDSEY, RW ;
CLARK, CR ;
LIEPONIS, J ;
MURPHY, MJ .
SPINE, 1989, 14 (10) :1051-1053
[7]
SONNTAG VKH, 1991, CLIN NEUR, V37, P636
[8]
A BIOMECHANICAL EVALUATION OF CERVICAL SPINAL STABILIZATION METHODS IN A BOVINE MODEL - STATIC AND CYCLICAL LOADING [J].
SUTTERLIN, CE ;
MCAFEE, PC ;
WARDEN, KE ;
REY, RM ;
FAREY, ID .
SPINE, 1988, 13 (07) :795-802
[9]
BIOMECHANICAL COMPARISON OF ANTERIOR CASPAR PLATE AND 3-LEVEL POSTERIOR FIXATION TECHNIQUES IN A HUMAN CADAVERIC MODEL [J].
TRAYNELIS, VC ;
DONAHER, PA ;
ROACH, RM ;
KOJIMOTO, H ;
GOEL, VK .
JOURNAL OF NEUROSURGERY, 1993, 79 (01) :96-103
[10]
ESOPHAGEAL PENETRATION BY AN ANTERIOR CERVICAL FIXATION DEVICE - A CASE-REPORT [J].
YEE, GKH ;
TERRY, AF .
SPINE, 1993, 18 (04) :522-527