THE USE OF A PORTABLE REFRACTOMETER TO MEASURE DESFLURANE

被引:3
作者
CHEAM, EWS
LOCKWOOD, GG
机构
[1] Department of Anaesthesia, Royal Postgraduate Medical School, Hammersmith Hospital, London W12 0HS, Du Cane Road
关键词
MEASUREMENT TECHNIQUES; REFRACTOMETER; INFRARED ABSORPTION; ANESTHETICS; VOLATILE; DESFLURANE;
D O I
10.1111/j.1365-2044.1995.tb15113.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
The measurement of volatile anaesthetic agents by infrared absorption is liable to interference by other gases which absorb infrared light. In particular, nitrous oxide may, in principle, interfere with the measurement of desflurane. On the other hand, the refractometer should be immune to such effects. We demonstrated that interactions may occur between desflurane and nitrous oxide with some gas analysers and that the refractometer remains a useful instrument.
引用
收藏
页码:607 / 610
页数:4
相关论文
共 4 条
[1]   PRECISE DETERMINATION OF REFRACTOMETRIC PARAMETERS FOR ANESTHETIC AGENT VAPORS [J].
ALLISON, JM ;
BIRCH, KP ;
CROWDER, JG .
APPLIED OPTICS, 1994, 33 (13) :2459-2464
[2]   PRECISE DETERMINATION OF REFRACTOMETRIC PARAMETERS FOR ATMOSPHERIC GASES [J].
BIRCH, KP .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1991, 8 (04) :647-651
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]   PORTABLE INTERFERENCE REFRACTOMETERS IN ANAESTHESIA [J].
HULANDS, GH ;
NUNN, JF .
BRITISH JOURNAL OF ANAESTHESIA, 1970, 42 (12) :1051-&