APPLICATION OF PARTICLE-BEAM MASS-SPECTROMETRY TO DRUGS - AN EXAMINATION OF THE PARAMETERS AFFECTING SENSITIVITY

被引:5
作者
TILLER, PR
机构
[1] Structural Chemistry Department, Glaxo Group Research Ltd., Greenford, Middlesex UB6 OHE, Greenford Road
来源
JOURNAL OF CHROMATOGRAPHY | 1993年 / 647卷 / 01期
关键词
D O I
10.1016/0021-9673(93)83329-Q
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The qualitative performance of a particle-beam interface is evaluated by an examination of the effects of altering the pressure of the nebulisation gas, the nebuliser position, the desolvation chamber temperature, the temperature of the nebulisation gas, the temperature of the source and the solvent composition. All these parameters have an impact on performance excepting the nebulisation gas temperature. A summary of the ''optimum'' settings for routine operation is given.