INVESTIGATION OF HOW DIFFERENT HALOS INFLUENCE PIN FORCES

被引:11
作者
KERWIN, GA [1 ]
CHOU, KL [1 ]
WHITE, DB [1 ]
SHEN, KL [1 ]
SALCICCIOLI, GG [1 ]
YANG, KH [1 ]
机构
[1] WAYNE STATE UNIV,CTR BIOENGN,DEPT ORTHOPAED SURG,818 W HANCOCK,DETROIT,MI 48202
关键词
Carvical spine fixation; Halo pin force;
D O I
10.1097/00007632-199405000-00016
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design. This study investigated the effect of the size and material of halo rings on the pin force produced at the halo pin-calvarium complex. Summary of Background Data: Loosening of halo pins has been a problem with using the halo-vest apparatus to stabilize cervical spine fractures. Methods. One standard halo pin and seven different halo rings were used to test the anterior pin hole of each ring at a constant torque of 8 inch-pounds (0.9 Nm). Results. Pin force fluctuated from halo to halo, and decreased as the nut was tightened. Conclusions. Identical torques applied to different halos will not guarantee the same magnitude of pin force for halo fixation.
引用
收藏
页码:1078 / 1081
页数:4
相关论文
共 10 条
[1]
Botte M.J., Byrne T.P., Garfin S.R., Application of the halo device for immobilization of the cervical spine utilizing an increased torque pressure, J Bone Joint Surg [Am], 69, pp. 750-752, (1987)
[2]
Botte M.J., Garfin S.R., Byrne T.P., Woo S.L.-Y., Nickel V.L., The halo skeletal fixator. Principles of application and maintenance, Clin Orthop, 239, pp. 12-18, (1989)
[3]
Garfin S.R., Botte M.J., Triggs K.J., Nickel V.L., Subdural abscess associated with halo-pin traction, J Bone Joint Surg [Am], 70, pp. 1338-1340, (1988)
[4]
Garfin S.R., Lee T.Q., Roux R.D., Et al., Structural behavior of the halo orthosis pin-bone interface: Biomechanical evaluation of standard and newly designed stainless steel halo fixation pins, Spine, 11, pp. 977-981, (1986)
[5]
Letts M., Kaylor D., Gouw G., A biomechanical analysis of halo fixation in children, J Bone Joint Surg [Br], 70, pp. 277-279, (1988)
[6]
Lind B., Sihlbom H., Nordwall A., Halo-vest treatment of unstable traumatic cervical spine injuries, Spine, 13, pp. 425-432, (1988)
[7]
Oberg E., Jones F.D., Horton H.L., Machinery’s Handbook, (1975)
[8]
Perry J., Nickel V.L., Total cervical spine fusion for neck paralysis, J Bone Joint Surg [Am], 41, pp. 37-59, (1959)
[9]
Triggs K.J., Ballock R.T., Lee T.Q., Woo S.L.-Y., Garfin S.R., The effect of angled insertion on halo pin fixation, Spine, 14, pp. 781-783, (1989)
[10]
Whitesides T.E., Mehserle W.L., Hutton W.C., The force exerted by the halo pin: A study comparing different halo systems, Spine, 17, (1992)