ON THE SPATIAL-RESOLUTION OF CLINICAL THERMOMETERS

被引:9
作者
GERIG, LH
SZANTO, J
RAAPHORST, GP
机构
[1] UNIV OTTAWA,FAC HLTH SCI,OTTAWA K1H 8MS,ONTARIO,CANADA
[2] CARLETON UNIV,DEPT PHYS,OTTAWA K1S 5B6,ONTARIO,CANADA
[3] UNIV OTTAWA,FAC HLTH SCI,OTTAWA K1H 8M5,ONTARIO,CANADA
关键词
HYPERTHERMIA; THERMOMETRY; THERMOCOUPLE;
D O I
10.1118/1.596897
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In this work the spatial resolution of type T (copper-Constantan) and type K (Chromel-Alumel) multipoint thermocouple thermometers was investigated. For clinical use, thermocouples are usually inserted within catheters to provide easy access, reproducibility, and a sterile environment. The effect of several types of catheter on the ability of these thermometers to accurately resolve thermal gradients was studied. The influence of the number of wire pairs within the multipoint thermometer was also investigated. A mathematical model has been developed to determine the spatial resolving power of these thermometers. Results indicate that type K thermocouples are generally superior to type T and the use of catheters plays a significant role in spatial resolution. Moreover, the use of mineral oil as a thermal coupling agent between the catheter and the thermometer was also found to have some effect.
引用
收藏
页码:679 / 684
页数:6
相关论文
共 10 条
  • [1] Cetas TC., 1990, INTRO PRACTICAL ASPE, P423
  • [2] DICKINSON RJ, 1985, PHYS MED BIOL, V13, P445
  • [3] GERIG LH, 1988, ENDOCURIETHER HYPERT, V4, P31
  • [4] THERMAL MAPPING IN EXPERIMENTAL CANCER-TREATMENT WITH HYPERTHERMIA - DESCRIPTION AND USE OF A SEMI-AUTOMATIC SYSTEM
    GIBBS, FA
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1983, 9 (07): : 1057 - 1063
  • [5] TEMPERATURE-MEASUREMENTS IN HIGH THERMAL-GRADIENTS .1. THE EFFECTS OF CONDUCTION
    LYONS, BE
    SAMULSKI, TV
    BRITT, RH
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1985, 11 (05): : 951 - 962
  • [6] MARTIN CJ, 1986, HYPERTHERMIA, P154
  • [7] OVERGAARD J, 1990, INTRO PRACTICAL ASPE, P213
  • [8] TEMPERATURE-MEASUREMENTS IN HIGH THERMAL-GRADIENTS .2. ANALYSIS OF CONDUCTION EFFECTS
    SAMULSKI, TV
    LYONS, BE
    BRITT, RH
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1985, 11 (05): : 963 - 971
  • [9] ERROR IN TEMPERATURE-MEASUREMENTS DUE TO CONDUCTION ALONG SENSOR LEADS
    SINGH, BS
    DYBBS, A
    [J]. JOURNAL OF HEAT TRANSFER-TRANSACTIONS OF THE ASME, 1976, 98 (03): : 491 - 495
  • [10] 1987, ASTM TECH PUB B, V470