CAPNOMETRY AND ANESTHESIA

被引:145
作者
BHAVANISHANKAR, K
MOSELEY, H
KUMAR, AY
DELPH, Y
机构
[1] Department of Anaesthesia, Queen Elizabeth Hospital, University of West Indies
来源
CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE | 1992年 / 39卷 / 06期
关键词
CARBON DIOXIDE; END-TIDAL; MEASUREMENT; MONITORING; MEASUREMENT TECHNIQUES; CAPNOMETRY;
D O I
10.1007/BF03008330
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
In the last decade, capnography has developed from a research instrument into a monitoring device considered to be essential during anaesthesia to ensure patient safety. Hence, a comprehensive understanding of capnography has become mandatory for the anaesthetist in charge of patients in the operating room and in the intensive care unit. This review of capnography includes the methods available to determine carbon dioxide in expired air, and an analysis of the physiology of capnograms, which are followed by a description of the applications of capnography in clinical practice. The theoretical backgrounds of the effect of barometric pressure, water vapour, nitrous oxide and other factors introducing errors in the accuracy of CO2 determination by the infra-red technique, currently the most popular method in use, are detailed. Physiological factors leading to changes in end-tidal carbon dioxide are discussed together with the clinical uses of this measurement to assess pulmonary blood flow indirectly, carbon dioxide production and adequacy of alveolar ventilation. The importance of understanding the shape of the capnogram as well as end-tidal carbon dioxide measurements is emphasized and its use in the early diagnosis of adverse events such as circuit disconnections, oesophageal intubation, defective breathing systems and hypoventilation is highlighted. Finally the precautions required in the use and interpretation of capnography are presented with the caveat that although no instrument will replace the continuous presence of the attentive physician, end-tidal carbon dioxide monitoring can be effective in the early detection of anaesthesia-related intraoperative accidents.
引用
收藏
页码:617 / 632
页数:16
相关论文
共 81 条
[1]  
Adams AP, 1989, RECENT ADV ANAESTHES, P155
[3]   END-TIDAL CO-2 MONITORING - ITS USE IN THE DIAGNOSIS AND MANAGEMENT OF MALIGNANT HYPERTHERMIA [J].
BAUDENDISTEL, L ;
GOUDSOUZIAN, N ;
COTE, C ;
STRAFFORD, M .
ANAESTHESIA, 1984, 39 (10) :1000-1003
[4]   CAPNOGRAPHIC DETECTION OF ANESTHESIA CIRCLE VALVE MALFUNCTIONS [J].
BERMAN, LS ;
PYLES, ST .
CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 1988, 35 (05) :473-475
[5]  
BIRMINGHAM PK, 1986, ANESTH ANALG, V65, P886
[6]   EFFECT OF PEEP ON THE ARTERIAL MINUS END-TIDAL CARBON-DIOXIDE GRADIENT [J].
BLANCH, L ;
FERNANDEZ, R ;
BENITO, S ;
MANCEBO, J ;
NET, A .
CHEST, 1987, 92 (03) :451-454
[7]   ACCURATE DETERMINATION OF END-TIDAL CARBON-DIOXIDE DURING ADMINISTRATION OF OXYGEN BY NASAL CANNULAE [J].
BOWE, EA ;
BOYSEN, PG ;
BROOME, JA ;
KLEIN, EF .
JOURNAL OF CLINICAL MONITORING, 1989, 5 (02) :105-110
[8]   ARTERIAL TO END-TIDAL CARBON-DIOXIDE TENSION DIFFERENCE DURING LAPAROSCOPY - MAGNITUDE AND EFFECT OF ANESTHETIC TECHNIQUE [J].
BRAMPTON, WJ ;
WATSON, RJ .
ANAESTHESIA, 1990, 45 (03) :210-214
[9]   IS THE TIDE TURNING [J].
BROWN, FN ;
MCINTYRE, RW .
CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 1990, 37 (01) :4-6
[10]   HOW THE RISE TIME OF CARBON-DIOXIDE ANALYZERS INFLUENCES THE ACCURACY OF CARBON-DIOXIDE MEASUREMENTS [J].
BRUNNER, JX ;
WESTENSKOW, DR .
BRITISH JOURNAL OF ANAESTHESIA, 1988, 61 (05) :628-638