CONTRAST-TO-NOISE-RATIO MEASUREMENTS IN 3-DIMENSIONAL MAGNETIC-RESONANCE ANGIOGRAPHY

被引:20
作者
DU, YP
PARKER, DL
DAVIS, WL
BLATTER, DD
机构
[1] UNIV UTAH,MED CTR,DEPT RADIOL,SALT LAKE CITY,UT 84132
[2] LATTER DAY ST HOSP,DEPT RADIOL,SALT LAKE CITY,UT 84143
[3] UNIV UTAH,DEPT PHYS,SALT LAKE CITY,UT 84112
关键词
MAGNETIC RESONANCE IMAGING; MAGNETIC RESONANCE ANGIOGRAPHY; CONTRAST-TO-NOISE RATIO;
D O I
10.1097/00004424-199311000-00005
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
RATIONALE AND OBJECTIVES. The authors report on the development and preliminary validation of a technique for measuring contrast-to-noise ratio (CNR) at all points along selected vessel segments in the original three-dimensional magnetic resonance angiography (MRA) dataset. METHODS. Contrast-to-noise ratio dependencies on flow rate, field of view, and flip angle were measured on images from a conventional time-of-flight MRA pulse sequence using constant flow in a branching vascular phantom. An estimate of the inherent variability of the technique was obtained from multiple scans of a flow phantom and a human volunteer. RESULTS. The overall standard deviation (SD) of the CNR was found to be approximately 6.1% of the average CNR value for the flow phantom study and 7.3% for the human study. Vessel CNR was found to increase with field of view and was found to become nonuniform for low flow rate and/or high flip angles. CONCLUSION. In general, such CNR measurements allow the investigation of the mechanism of signal loss and general technique optimization in MRA.
引用
收藏
页码:1004 / 1009
页数:6