MAGNETIC RESONANCE IMAGING;
MAGNETIC RESONANCE ANGIOGRAPHY;
CONTRAST-TO-NOISE RATIO;
D O I:
10.1097/00004424-199311000-00005
中图分类号:
R8 [特种医学];
R445 [影像诊断学];
学科分类号:
1002 ;
100207 ;
1009 ;
摘要:
RATIONALE AND OBJECTIVES. The authors report on the development and preliminary validation of a technique for measuring contrast-to-noise ratio (CNR) at all points along selected vessel segments in the original three-dimensional magnetic resonance angiography (MRA) dataset. METHODS. Contrast-to-noise ratio dependencies on flow rate, field of view, and flip angle were measured on images from a conventional time-of-flight MRA pulse sequence using constant flow in a branching vascular phantom. An estimate of the inherent variability of the technique was obtained from multiple scans of a flow phantom and a human volunteer. RESULTS. The overall standard deviation (SD) of the CNR was found to be approximately 6.1% of the average CNR value for the flow phantom study and 7.3% for the human study. Vessel CNR was found to increase with field of view and was found to become nonuniform for low flow rate and/or high flip angles. CONCLUSION. In general, such CNR measurements allow the investigation of the mechanism of signal loss and general technique optimization in MRA.