CANINE MODEL OF ACUTE AORTIC RUPTURE - TREATMENT WITH PERCUTANEOUS DELIVERY OF A COVERED Z-STENT - WORK-IN-PROGRESS

被引:14
作者
WILLIAMS, DM [1 ]
ANDREWS, JC [1 ]
CHEE, SS [1 ]
MARX, MV [1 ]
ABRAMS, GD [1 ]
机构
[1] UNIV MICHIGAN HOSP,DEPT PATHOL,ANN ARBOR,MI 48109
关键词
D O I
10.1016/S1051-0443(94)71608-0
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To develop a percutaneous treatment for aortic rupture with use of a covered intraluminal stent. MATERIALS AND METHODS: A transmural tear was created percutaneously in the thoracic aorta in six dogs with use of a 4-mm angioplasty balloon, Gianturco Z stents were covered with polytetrafluoroethylene, loaded into a 14-F sheath, and advanced through the femoral artery to the site of injury. Within 2 minutes after initiation of the injury, the stent was deployed, Homologous canine blood was given during the procedure, Dogs that survived 24 hours were then killed. RESULTS: The first stent did not expand completely, and the dog died in 1 hour, At necropsy, the first two dogs (1-hour and 8-hour survival) had a large left hemothorax and extensive periaortic hematoma, indicating intrathoracic exsanguination. The next four dogs were treated with a modified stent and survived 2 hours (n = 1), 8 hours (n = 1), and 24 hours (n = 2), At necropsy hemothorax did not exceed 15 mL, and periaortic hematomas were small. The cause of death in the two early casualties with the modified stent is uncertain, There were no signs of spinal cord injury despite occlusion of three pairs of intercostal arteries. CONCLUSION: The covered Z stent (in its modified form) tamponaded the aortic tear, preventing exsanguination. Long-term studies of biocompatibility of this device appear justified.
引用
收藏
页码:797 / 803
页数:7
相关论文
共 14 条
[1]   TRANSFEMORAL ENDOVASCULAR AORTIC GRAFT PLACEMENT [J].
CHUTER, TAM ;
GREEN, RM ;
OURIEL, K ;
FIORE, WM ;
DEWEESE, JA ;
QUERAL, LA ;
CRIADO ;
NATH, RL ;
BERKOWITZ, HD ;
SUMPIO, BE .
JOURNAL OF VASCULAR SURGERY, 1993, 18 (02) :185-197
[2]   THE PENETRATING AORTIC ULCER - PATHOLOGIC MANIFESTATIONS, DIAGNOSIS, AND MANAGEMENT [J].
COOKE, JP ;
KAZMIER, FJ ;
ORSZULAK, TA .
MAYO CLINIC PROCEEDINGS, 1988, 63 (07) :718-725
[3]   ELASTIC CHARACTERISTICS OF THE SELF-EXPANDING METALLIC STENTS [J].
FALLONE, BG ;
WALLACE, S ;
GIANTURCO, C .
INVESTIGATIVE RADIOLOGY, 1988, 23 (05) :370-376
[4]   PENETRATING ATHEROSCLEROTIC ULCERS OF THE AORTA [J].
HARRIS, JA ;
BIS, KG ;
GLOVER, JL ;
BENDICK, PJ ;
SHETTY, A ;
BROWN, OW .
JOURNAL OF VASCULAR SURGERY, 1994, 19 (01) :90-99
[5]   PENETRATING ATHEROSCLEROTIC ULCERS OF THE THORACIC AORTA [J].
HUSSAIN, S ;
GLOVER, JL ;
BREE, R ;
BENDICK, PJ .
JOURNAL OF VASCULAR SURGERY, 1989, 9 (05) :710-717
[6]   RELOCATABLE GIANTURCO EXPANDABLE METALLIC STENTS [J].
IRIE, T ;
FURUI, S ;
YAMAUCHI, T ;
MAKITA, K ;
SAWADA, S ;
TAKENAKA, E .
RADIOLOGY, 1991, 178 (02) :575-578
[7]   TREATMENT OF ACUTE TRAUMATIC RUPTURE OF AORTA - 10-YEAR EXPERIENCE [J].
KIRSH, MM ;
BEHRENDT, DM ;
ORRINGER, MB ;
GAGO, O ;
GRAY, LA ;
MILLS, LJ ;
WALTER, JF ;
SLOAN, H .
ANNALS OF SURGERY, 1976, 184 (03) :308-316
[8]   INTRALUMINAL BYPASS OF ABDOMINAL AORTIC-ANEURYSM - FEASIBILITY STUDY [J].
LABORDE, JC ;
PARODI, JC ;
CLEM, MF ;
TIO, FO ;
BARONE, HD ;
RIVERA, FJ ;
ENCARNACION, CE ;
PALMAZ, JC .
RADIOLOGY, 1992, 184 (01) :185-190
[9]  
MURRAY CA, 1973, CIRCULATION, V47, P848, DOI 10.1161/01.CIR.47.4.848
[10]   NONPENETRATING TRAUMATIC INJURY OF THE AORTA [J].
PARMLEY, LF ;
MATTINGLY, TW ;
MANION, WC ;
JAHNKE, EJ .
CIRCULATION, 1958, 17 (06) :1086-1101