Comparative analysis between 5 mm and 7.5 mm collimators in CyberKnife radiosurgery for trigeminal neuralgia

被引:11
作者
Sudahar, H. [1 ]
Kurup, P. G. G. [1 ]
Murali, V. [1 ]
Velmurugan, J. [2 ]
机构
[1] Apollo Special Hosp, Dept Radiotherapy, 320 Padma Complex,Mt Rd, Madras 600035, Tamil Nadu, India
[2] Anna Univ, Dept Med Phys, Madras, Tamil Nadu, India
关键词
Collimators; CyberKnife; stereotactic radiosurgery; trigeminal neuralgia;
D O I
10.4103/0971-6203.116364
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 [临床医学]; 100207 [影像医学与核医学]; 1009 [特种医学];
摘要
Trigeminal neuralgia (TN) is treated in CyberKnife (Accuray Inc, Sunnyvale, USA) with the 5 mm collimator whose dosimetric inaccuracy is higher than the other available collimators. The 7.5 mm collimator which is having less dosimetric uncertainty can be an alternative for 5 mm collimator provided the dose distribution with 7.5 mm collimator is acceptable. Aim of this study is to analyze the role of 7.5 mm collimator in CyberKnife treatment plans of TN. The treatment plans with 5 mm collimators were re-optimized with 7.5 mm collimator and a bi-collimator system (5 mm and 7.5 mm). The treatment plans were compared for target coverage, brainstem doses, and the dose to normal tissues. The target and brainstem doses were comparable. However, the conformity indices were 2.31 +/- 0.52, 2.40 +/- 0.87 and 2.82 +/- 0.51 for 5 mm, bi-collimator (5mm and 7.5 mm), 7.5 mm collimator plans respectively. This shows the level of dose spillage in 7.5 mm collimator plans. The 6 Gy dose volumes in 7.5 mm plans were 1.53 and 1.34 times higher than the 5 mm plan and the bi-collimator plans respectively. The treatment time parameters were lesser for 7.5 mm collimators. Since, the normal tissue dose is pretty high in 7.5 mm collimator plans, the use of it in TN plans can be ruled out though the treatment time is lesser for these 7.5 mm collimator plans.
引用
收藏
页码:120 / 124
页数:5
相关论文
共 28 条
[1]
Agrawal Shaila M, 2010, Natl J Maxillofac Surg, V1, P179, DOI 10.4103/0975-5950.79226
[2]
Outcome for Patients with Essential Trigeminal Neuralgia Treated with Linear Accelerator Stereotactic Radiosurgery [J].
Antonio dos Santos, Marcos ;
Bustos Perez de Salcedo, Jose ;
Gutierrez Diaz, Jose Angel ;
Nagore, Gorka ;
Calvo, Felipe A. ;
Samblas, Jose ;
Marsiglia, Hugo ;
Sallabanda, Kita .
STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2011, 89 (04) :220-225
[3]
Brisman R, 2012, WORLD NEUROSURG, V78, DOI [10.1016/J.WNEU.2012.02.043, 10.1016/j.wneu.2012.02.043]
[4]
Prospective Study of the Short-Term Adverse Effects of Gamma Knife Radiosurgery [J].
Chao, Samuel T. ;
Thakkar, Vipul V. ;
Barnett, Gene H. ;
Vogelbaum, Michael A. ;
Angelov, Lilyana ;
Weil, Robert J. ;
Rasmussen, Peter ;
Reuther, Alwyn M. ;
Jamison, Betty ;
Neyman, Gennady ;
Suh, John H. .
TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2012, 11 (02) :117-122
[5]
Does increased nerve length within the treatment volume improve trigeminal neuralgia radiosurgery? A prospective double-blind, randomized study [J].
Flickinger, JC ;
Pollock, BE ;
Kondziolka, D ;
Phuong, LK ;
Foote, RL ;
Stafford, SL ;
Lunsford, LD .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 51 (02) :449-454
[6]
Preliminary results of 45 patients with trigeminal neuralgia treated with radiosurgery compared to hypofractionated stereotactic radiotherapy, using a dedicated linear accelerator [J].
Fraioli, Mario Francesco ;
Strigari, Lidia ;
Fraioli, Chiara ;
Lecce, Mario ;
Lisciani, Damiano .
JOURNAL OF CLINICAL NEUROSCIENCE, 2012, 19 (10) :1401-1403
[7]
Monte Carlo simulated correction factors for machine specific reference field dose calibration and output factor measurement using fixed and iris collimators on the CyberKnife system [J].
Francescon, P. ;
Kilby, W. ;
Satariano, N. ;
Cora, S. .
PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (12) :3741-3758
[8]
Gerbi Bruce J, 2004, J Appl Clin Med Phys, V5, P80, DOI 10.1120/jacmp.2021.25278
[9]
Gorgulho Alessandra, 2012, Surg Neurol Int, V3, pS17, DOI 10.4103/2152-7806.91606
[10]
Gamma knife radiosurgery for trigeminal neuralgia: Comparing the use of a 4-mm versus concentric 4- and 8-mm collimators [J].
Kanner, AA ;
Neyman, G ;
Suh, JH ;
Weinhous, MS ;
Lee, SY ;
Barnett, GH .
STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2004, 82 (01) :49-57