KNOWLEDGE-BASED COMPUTER-SYSTEM FOR THE DETECTION OF MATRIX INTERFERENCES IN ATOMIC-ABSORPTION SPECTROMETRIC METHODS

被引:8
作者
PENNINCKX, W
VANKEERBERGHEN, P
MASSART, DL
SMEYERSVERBEKE, J
机构
[1] ChemoAC, Pharmaceutical Institute, Vrije Universiteit Brüssel, 1090 Brussels
关键词
ATOMIC ABSORPTION SPECTROMETRY; METHOD VALIDATION; MATRIX INTERFERENCES; COMPUTER SYSTEMS;
D O I
10.1039/ja9951000207
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A knowledge-based computer system which assists with the detection of matrix interferences during the validation of atomic absorption spectrometry (AAS) is described. The system compares the slopes of a standard addition line and an aqueous calibration line. An experimental design is proposed which guarantees that the probability of not detecting important interferences (beta-error) is acceptably low. To permit a detection of interferences by the comparison of curved calibration lines, a linearization is proposed. The system is developed in a Windows environment add is one module of a more complete hypertext system for the final validation of AAS methods.
引用
收藏
页码:207 / 214
页数:8
相关论文
共 13 条
[1]  
COOPER BE, STATISTICS EXPT, P69
[2]  
De Galan L, 1969, SPECTROCHIM ACTA B, V24, P679
[3]   DETERMINATION OF STRONGLY CURVED CALIBRATION GRAPHS IN FLAME ATOMIC-ABSORPTION SPECTROMETRY - COMPARISON OF MANUALLY DRAWN AND COMPUTER-CALCULATED GRAPHS [J].
DEGALAN, L ;
VANDALEN, HPJ ;
KORNBLUM, GR .
ANALYST, 1985, 110 (04) :323-329
[4]   LINEARIZATION OF CALIBRATION CURVES IN ZEEMAN ATOMIC-ABSORPTION SPECTROMETRY [J].
LVOV, BV ;
POLZIK, LK ;
KOCHAROVA, NV ;
NEMETS, YA ;
NOVICHIKHIN, AV .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1992, 47 (10) :1187-1202
[5]   THEORETICAL-ANALYSIS OF CALIBRATION CURVES FOR GRAPHITE-FURNACE ATOMIC-ABSORPTION SPECTROMETRY [J].
LVOV, BV ;
POLZIK, LK ;
KOCHAROVA, NV .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1992, 47 (07) :889-895
[6]  
Massart D. L., 1978, EVALUATION OPTIMIZAT
[7]  
Massart D. L., 1988, CHEMOMETRICS TXB
[8]   METHOD VALIDATION - SOFTWARE TO COMPARE THE SLOPES OF 2 CALIBRATION LINES WITH DIFFERENT RESIDUAL VARIANCE [J].
MASSART, DL ;
SMEYERSVERBEKE, J ;
RIUS, FX .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 1989, 8 (02) :49-51
[9]  
MASSART DL, 1988, TRENDS ANAL CHEM, V8, P8
[10]  
PENNINCKX W, 1994, ANALUSIS, V22, pM22