PERCUTANEOUS TRANSLUMINAL CORONARY ROTARY ABLATION WITH ROTABLATOR (EUROPEAN EXPERIENCE)

被引:88
作者
BERTRAND, ME
LABLANCHE, JM
LEROY, F
BAUTERS, C
DEJAEGERE, P
SERRUYS, PW
MEYER, J
DIETZ, U
ERBEL, R
机构
[1] UNIV LILLE,HOP CARDIOL,DIV CARDIOL B,LILLE,FRANCE
[2] THORAX CTR ROTTERDAM,CATHETERISAT LAB,ROTTERDAM,NETHERLANDS
[3] UNIV MAINZ,MED CLIN 2,W-6500 MAINZ,GERMANY
关键词
D O I
10.1016/0002-9149(92)90988-B
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
This study reports the results from 3 European centers using rotary ablation with Rotablator(TM), a device that is inserted into the coronary artery and removes atheroma by grinding it into millions of tiny fragments. Rotary ablation was performed in 129 patients. Primary success (reduction in percent luminal narrowing > 20%, residual stenosis < 50%, without complications) was achieved by rotary angioplasty alone in 73 patients (57%). An additional 38 patients (29%) had successful adjunctive balloon angioplasty. Thus primary success was achieved in 111 patients (86%) at the end of the procedure. Acute occlusion occurred in 10 patients (7.7%). Recanalization was achieved by balloon angioplasty in 7: urgent bypass grafting was undertaken in 2. Q-wave and non-Q-wave myocardial infarction occurred in 3 and 7 patients, respectively. No deaths occurred. Following angiography was performed in 74 patients (60%). Restenosis, defined as the recurrence of significant luminal narrowing (> 50%) occurred in 17 of 37 patients (46%) who underwent rotary ablation alone, and 11 of 37 patients (30%) who had adjunctive balloon angioplasty. The overall angiographic restenosis rate was 37.8%. In conclusion, rotary ablation is technically feasible, and relatively safe in the coronary circulation. The low primary success rate reflects the limited size of the device, which can be introduced through available guiding catheters, and limits the use of rotary ablation as a stand-alone procedure to lesions in small arteries or in distal locations. No reduction in restenosis was seen, but the role of this device combined with balloon angioplasty in larger arteries needs to be further defined.
引用
收藏
页码:470 / 474
页数:5
相关论文
共 8 条
  • [1] REMOVAL OF FOCAL ATHEROMATOUS LESIONS BY ANGIOSCOPICALLY GUIDED HIGH-SPEED ROTARY ATHERECTOMY - PRELIMINARY EXPERIMENTAL-OBSERVATIONS
    AHN, SS
    AUTH, D
    MARCUS, DR
    MOORE, WS
    [J]. JOURNAL OF VASCULAR SURGERY, 1988, 7 (02) : 292 - 300
  • [2] BERTRAND ME, 1990, HERZ, V15, P285
  • [3] BERTRAND ME, 1990, TXB INTERVENTIONAL C, P580
  • [4] PERCUTANEOUS CORONARY ROTATIONAL ANGIOPLASTY IN HUMANS - PRELIMINARY-REPORT
    FOURRIER, JL
    BERTRAND, ME
    AUTH, DC
    LABLANCHE, JM
    GOMMEAUX, A
    BRUNETAUD, JM
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1989, 14 (05) : 1278 - 1282
  • [5] GRUENTZIG AR, 1979, NEW ENGL J MED, V301, P61
  • [6] ROTATIONAL ATHERECTOMY IN ATHEROSCLEROTIC RABBIT ILIAC ARTERIES
    HANSEN, DD
    AUTH, DC
    VRACKO, R
    RITCHIE, JL
    [J]. AMERICAN HEART JOURNAL, 1988, 115 (01) : 160 - 165
  • [7] RESTENOSIS AFTER PERCUTANEOUS TRANS-LUMINAL CORONARY ANGIOPLASTY (PTCA) - A REPORT FROM THE PTCA REGISTRY OF THE NATIONAL-HEART-LUNG-AND-BLOOD-INSTITUTE
    HOLMES, DR
    VLIETSTRA, RE
    SMITH, HC
    VETROVEC, GW
    KENT, KM
    COWLEY, MJ
    FAXON, DP
    GRUENTZIG, AR
    KELSEY, SF
    DETRE, KM
    VANRADEN, MJ
    MOCK, MB
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1984, 53 (12) : C77 - C81
  • [8] ANALYSIS OF HIGH-FREQUENCY ROTATIONAL ANGIOPLASTY-INDUCED ECHO CONTRAST
    ZOTZ, R
    STAHR, P
    ERBEL, R
    AUTH, D
    MEYER, J
    [J]. CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1991, 22 (02): : 137 - 144