The analysis of nitric oxide-derived nitrite and nitrate ions in biological fluids represents a proven strategy for determining nitric oxide participation in a diverse range of physiological and pathophysiological processes in vivo. In this article we describe a versatile method for the simultaneous measurement of NO2- and NO3- anions in both plasma and isolated tumour models based on anion-exchange chromatography with spectrophotometric detection (214 nm). This method compares well with the capillary electrophoresis technique, exhibiting an equivalent sensitivity for NO2-/NO3- anions and short run-times, i.e. not greater than 4 min. Comparisons are also made with two alternative but less satisfactory methods which employ ion-exchange or reversed-phase ion-pair chromatography with conductimetric as well as spectrophotometric detection. Technical problems associated with each method, particularly those arising from nitrate contamination, have been addressed.