HEATING PATTERN OF HELICAL MICROWAVE INTRACAVITARY ESOPHAGEAL APPLICATOR

被引:10
作者
LIU, RL [1 ]
ZHANG, EY [1 ]
GROSS, EJ [1 ]
CETAS, TC [1 ]
机构
[1] ARIZONA HLTH SCI CTR,DIV RADIAT ONCOL,TUCSON,AZ 85724
关键词
433 MHZ AND 915 MHZ HMIO APPLICATORS; THERMOGRAM; SAR; HEATING PATTERN; INTRACAVITARY HYPERTHERMIA;
D O I
10.3109/02656739109034970
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Helical microwave intracavitary oesophageal (HMIO) applicators were designed to operate at frequencies of 433 MHz and 915 MHz. Heating patterns were studied within muscle-equivalent phantom by thermographic camera and fibreoptic thermometers. The results showed that frequency significantly influenced the microwave heating pattern. The 433 MHz applicator had a single power deposition region, the longitudinal specific absorption rate (SAR) distribution appeared to be nearly even, and the maximum SAR value occurred close to the centre of the active length of the applicator. The 915 MHz applicator had two power deposition regions, the peak SAR values occurred at about 1/4 and 3/4 of the active length respectively. The radial SAR distribution suggested that there is no obvious difference between the 433 MHz and 915 MHz applicators in that the average radial penetration of 50% surface SAR (RP50) was about 0.65 cm. It was also shown that power deposition was axially symmetric for both 433 MHz and 915 MHz HMIO applicators. It is shown that better impedance matching is more important for intracavitary hyperthermia than for external hyperthermia. Choosing HMIO applicators in clinical practice is also discussed.
引用
收藏
页码:577 / 586
页数:10
相关论文
共 20 条
[1]   MICROWAVE APPLICATOR FOR TRANS-URETHRAL HYPERTHERMIA OF BENIGN PROSTATIC HYPERPLASIA [J].
ASTRAHAN, MA ;
SAPOZINK, MD ;
COHEN, D ;
LUXTON, G ;
KAMPP, TD ;
BOYD, S ;
PETROVICH, Z .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1989, 5 (03) :283-296
[2]   A TECHNIQUE FOR COMBINING MICROWAVE HYPERTHERMIA WITH INTRALUMINAL BRACHYTHERAPY OF THE ESOPHAGUS [J].
ASTRAHAN, MA ;
SAPOZINK, MD ;
LUXTON, G ;
KAMPP, TD ;
PETROVICH, Z .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1989, 5 (01) :37-51
[3]  
BASSEN HI, 1982, PHYSICAL ASPECTS HYP, P372
[4]  
CETAS T, 1982, PHYS ASPECTS HYPERTH, P441
[5]   ERROR SOURCES AFFECTING THERMOCOUPLE THERMOMETRY IN RF ELECTROMAGNETIC-FIELDS [J].
CHAKRABORTY, DP ;
BREZOVICH, IA .
JOURNAL OF MICROWAVE POWER AND ELECTROMAGNETIC ENERGY, 1982, 17 (01) :17-28
[6]  
CHOU CK, 1984, BIOELECTROMAGNETICS, V5, P435, DOI 10.1002/bem.2250050408
[8]   INCREASED CURES USING FRACTIONATED EXPOSURES OF X-IRRADIATION AND HYPERTHERMIA IN LOCAL TREATMENT OF RIDGWAY OSTEOGENIC SARCOMA IN MICE [J].
HAHN, EW ;
ALFIERI, AA ;
KIM, JH .
RADIOLOGY, 1974, 113 (01) :199-202
[9]  
HOU BS, 1989, CANCER, V64, P1777, DOI 10.1002/1097-0142(19891101)64:9<1777::AID-CNCR2820640903>3.0.CO
[10]  
2-I