LABORATORY VALIDATION-STUDY OF A PORTABLE SYSTEM FOR REMOTE RECORDING OF SLEEP-RELATED RESPIRATORY DISORDERS

被引:35
作者
ORR, WC
EIKEN, T
PEGRAM, V
JONES, R
RUNDELL, OH
机构
[1] BAPTIST MONTCLAIR MED CTR,BIRMINGHAM,AL
[2] BAPTIST MED CTR,OKLAHOMA CITY,OK 73112
关键词
D O I
10.1378/chest.105.1.160
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
This study represents the validation of an out-of-laboratory recording system for screening sleep disorders by comparing it to an in-laboratory system which accomplishes complete polysomnography. Forty patient cases were collected. Each patient slept in the sleep laboratory with the simultaneous monitoring of eight channels from the Sleep I/T (CNS, Inc), as well as complete polysomnographic monitoring comparable to that which accomplishes routine patient evaluations utilizing the Sleeplab (CNS, Inc) system. The data collected on the Sleeplab system were analyzed in a computer-assisted manner followed by manual overreading of the raw data. The Sleep I/T data analysis was performed automatically using default parameters. The results revealed no significant differences in the determination of sleep efficiency, respiratory disturbance index (RDI), periodic leg movement (PLM) index, or the desaturation index. For the RDI, the sensitivity was 100 percent and the specificity was 93 percent for correctly detecting an RDI of 15 per hour, and the PLM index had a sensitivity of 89 percent and specificity of 97 percent in detecting episodes greater than 20 per hour. These data demonstrate excellent agreement between this out-of-laboratory system and polysomnography in its ability to detect accurately RDI, PLM index, and saturation levels.
引用
收藏
页码:160 / 162
页数:3
相关论文
共 4 条
  • [1] CLINICAL-VALUE OF POLYSOMNOGRAPHY
    DOUGLAS, NJ
    THOMAS, S
    JAN, MA
    [J]. LANCET, 1992, 339 (8789) : 347 - 350
  • [2] MEASUREMENT OF SLEEP-RELATED BREATHING DISTURBANCES IN EPIDEMIOLOGIC STUDIES - ASSESSMENT OF THE VALIDITY AND REPRODUCIBILITY OF A PORTABLE MONITORING DEVICE
    REDLINE, S
    TOSTESON, T
    BOUCHER, MA
    MILLMAN, RP
    [J]. CHEST, 1991, 100 (05) : 1281 - 1286
  • [3] A LIMITED DIAGNOSTIC INVESTIGATION FOR OBSTRUCTIVE SLEEP-APNEA SYNDROME - OXIMETRY AND STATIC CHARGE SENSITIVE BED
    SVANBORG, E
    LARSSON, H
    CARLSSONNORDLANDER, B
    PIRSKANEN, R
    [J]. CHEST, 1990, 98 (06) : 1341 - 1345
  • [4] SCREENING FOR SLEEP-APNEA USING PULSE OXIMETRY AND A CLINICAL SCORE
    WILLIAMS, AJ
    YU, G
    SANTIAGO, S
    STEIN, M
    [J]. CHEST, 1991, 100 (03) : 631 - 635