HEAT AND MOISTURE EXCHANGER VS HEATED HUMIDIFIER DURING LONG-TERM MECHANICAL VENTILATION - A PROSPECTIVE RANDOMIZED STUDY

被引:63
作者
MISSET, B
ESCUDIER, B
RIVARA, D
LECLERCQ, B
NITENBERG, G
机构
关键词
D O I
10.1378/chest.100.1.160
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Adequate humidification of inspired gases with HMEs during long-term MV remains controversial. In this study, a comparison is made between tracheal secretions during long-term MV either with HME or conventional HH. Both the HME and HH groups were similar with respect to age, sex, diagnosis, duration of MV, SAPS and mortality. Temperature of gases in the tracheal tube was lower and the amount of tracheal instillations was greater in the HME group than in the HH group. Tracheal secretions became thicker between day 1 (control) and day 5, in the HME group than in the HH group. Four and two tube occlusions occurred in HME and HH groups, respectively. Tracheal bacterial colonization was similar in the two groups. Given the advantages of HME (reduced nurses' work and financial cost), HME could be routinely used under cautious surveillance and replaced by HH if difficulty in suctioning occurs.
引用
收藏
页码:160 / 163
页数:4
相关论文
共 17 条
[1]  
BOITEAU R, 1988, REAN SOINS INTENS ME, V4, P382
[2]  
BOITEAU R, 1989, REAN SOINS INTENS ME, V5, P459
[3]  
BOUILHAC M, 1984, Agressologie, V25, P299
[4]   HUMIDITY AND THE ANESTHETIZED PATIENT [J].
CHALON, J ;
PATEL, C ;
ALI, M ;
RAMANATHAN, S ;
CAPAN, L ;
TANG, CK ;
TURNDORF, H .
ANESTHESIOLOGY, 1979, 50 (03) :195-198
[5]  
CHALON J, 1984, ANESTH ANALG, V63, P566
[6]   ENDOTRACHEAL-TUBE OCCLUSION ASSOCIATED WITH THE USE OF HEAT AND MOISTURE EXCHANGERS IN THE INTENSIVE-CARE UNIT [J].
COHEN, IL ;
WEINBERG, PF ;
FEIN, IA ;
ROWINSKI, GS .
CRITICAL CARE MEDICINE, 1988, 16 (03) :277-279
[7]   CONTAMINATION OF MECHANICAL VENTILATORS WITH TUBING CHANGES EVERY 24 OR 48 HOURS [J].
CRAVEN, DE ;
CONNOLLY, MG ;
LICHTENBERG, DA ;
PRIMEAU, PJ ;
MCCABE, WR .
NEW ENGLAND JOURNAL OF MEDICINE, 1982, 306 (25) :1505-1509
[8]   EFFICACY OF A NEW HYGROSCOPIC CONDENSER HUMIDIFIER [J].
HAY, R ;
MILLER, WC .
CRITICAL CARE MEDICINE, 1982, 10 (01) :49-51
[9]  
MARTIN C, 1988, REAN SOINS INTENS ME, V4, P382
[10]  
MORITZ AR, 1945, ARCH INTERN MED, V15, P33