CAPILLARY ELECTROPHORESIS LASER FLUOROMETRY INSTRUMENTATION FOR THE FACILE OPTIMIZATION OF DNA SEPARATIONS USING IN-SITU SIZE-SELECTIVE GRADIENTS AND ADJUSTABLE DETECTION ZONE

被引:15
作者
CLARK, BK [1 ]
VO-DINH, T [1 ]
SEPANIAK, MJ [1 ]
机构
[1] UNIV TENNESSEE, DEPT CHEM, KNOXVILLE, TN 37996 USA
关键词
D O I
10.1021/ac00099a030
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The construction and evaluation of a capillary electrophoresis instrument for the facile optimization of DNA separations is described herein. This instrumentation allows for convenient translation of the detection zone with little loss in detectability. Changing the effective capillary length is shown to be important when complete resolution of specified DNA fragments in a sample, in the minimum time, is desired, Efficiency is found to decrease with column length in accordance with limits imposed by axial diffusion. However, complete resolution of the fragments in a Phi X-174-HeaIII digest is accomplished in 155 s at moderate held strengths and an effective capillary length of 15 cm. It is also demonstrated that in situ gradients can be established using this instrumentation. The general chromatographic elution problem is addressed with gradients, resulting in further improvements in separation performance relative to isocratic separations.
引用
收藏
页码:680 / 683
页数:4
相关论文
共 21 条
[1]   RAPID SEPARATION OF DNA RESTRICTION DIGESTS USING SIZE-SELECTIVE CAPILLARY ELECTROPHORESIS WITH APPLICATION TO DNA-FINGERPRINTING [J].
CLARK, BK ;
NICKLES, CL ;
MORTON, KC .
JOURNAL OF MICROCOLUMN SEPARATIONS, 1994, 6 (05) :503-513
[2]   EVALUATION OF ON-COLUMN LABELING WITH INTERCALATING DYES FOR FLUORESCENCE DETECTION OF DNA FRAGMENTS SEPARATED BY CAPILLARY ELECTROPHORESIS [J].
CLARK, BK ;
SEPANIAK, MJ .
JOURNAL OF MICROCOLUMN SEPARATIONS, 1993, 5 (03) :275-282
[3]   IMPROVED SEPARATION OF NUCLEIC-ACIDS WITH ANALYTE VELOCITY MODULATION CAPILLARY ELECTROPHORESIS [J].
DEMANA, T ;
LANAN, M ;
MORRIS, MD .
ANALYTICAL CHEMISTRY, 1991, 63 (23) :2795-2797
[4]   DIRECT EXAMINATION OF THE INJECTION PROCESS IN LIQUID-CHROMATOGRAPHIC SEPARATIONS [J].
EVANS, CE ;
MCGUFFIN, VL .
ANALYTICAL CHEMISTRY, 1991, 63 (14) :1393-1402
[5]  
Grossman P.D., 1992, CAPILLARY ELECTROPHO, P352
[6]   EFFECT OF TEMPERATURE ON THE SEPARATION OF DNA RESTRICTION FRAGMENTS IN CAPILLARY GEL-ELECTROPHORESIS [J].
GUTTMAN, A ;
COOKE, N .
JOURNAL OF CHROMATOGRAPHY, 1991, 559 (1-2) :285-294
[7]   EFFECT OF TOTAL PERCENT POLYACRYLAMIDE IN CAPILLARY GEL-ELECTROPHORESIS FOR DNA SEQUENCING OF SHORT FRAGMENTS - A PHENOMENOLOGICAL MODEL [J].
HARKE, HR ;
BAY, S ;
ZHANG, JZ ;
ROCHELEAU, MJ ;
DOVICHI, NJ .
JOURNAL OF CHROMATOGRAPHY, 1992, 608 (1-2) :143-150
[8]   HIGH-PERFORMANCE ELECTROPHORESIS - ELIMINATION OF ELECTROENDOSMOSIS AND SOLUTE ADSORPTION [J].
HJERTEN, S .
JOURNAL OF CHROMATOGRAPHY, 1985, 347 (02) :191-198
[9]   DNA SEQUENCING USING CAPILLARY ARRAY ELECTROPHORESIS [J].
HUANG, XHC ;
QUESADA, MA ;
MATHIES, RA .
ANALYTICAL CHEMISTRY, 1992, 64 (18) :2149-2154
[10]   HIGH-RESOLUTION LIQUID-CHROMATOGRAPHY OF DNA FRAGMENTS ON NONPOROUS POLY(STYRENE-DIVINYLBENZENE) PARTICLES [J].
HUBER, CG ;
OEFNER, PJ ;
PREUSS, E ;
BONN, GK .