AMPEROMETRIC DETECTION OF AMINO-ACIDS IN A FLOW-INJECTION SYSTEM WITH A NICKEL(II)-MODIFIED ELECTRODE WITH AN EASTMAN-AQ POLYMER FILM

被引:8
作者
LIU, AH [1 ]
WANG, EK [1 ]
机构
[1] CHINESE ACAD SCI,CHANGCHUN INST APPL CHEM,ELECTROANALYT CHEM LAB,CHANGCHUN 130022,PEOPLES R CHINA
关键词
AMPEROMETRY; FLOW INJECTION; LIQUID CHROMATOGRAPHY; AMINO ACIDS; CHEMICALLY MODIFIED ELECTRODES;
D O I
10.1016/0003-2670(93)85125-4
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Amperometic flow measurements were made at +0.55 V (vs. Ag/AgCl) in 0.1 mol l-1 KOH electrolyte with an Ni(II) chemically modified electrode (CME) with an Eastman-AQ polymer film. The use and characteristics of a Ni(II)-containing crystalline and polymer-modified electrode obtained by a double coating step as a detector for amino acids in a flow-injection system using reversed-phase liquid chromatography are described. The detection of these analytes is based on the higher oxidation state of nickel (NiOOH) controlled by the applied potential. The electroanalytical parameters and the detection current for a series of amines and amino acids were investigated. The use of such a CME in the flow-injection technique was found to be suitable in a solution at low pH. The linear range for glycine is 5 X 10(-6)-0.1 mol 1-1 with a detection limit of 1.0 X 10(-6) mol l-1. A 1 X 10(-4) mol 1-1 mixture of serine and tyrosine was also detected after separation on an Nucleosil C18 column.
引用
收藏
页码:223 / 229
页数:7
相关论文
共 23 条
[1]   CONTINUOUS-FLOW POTENTIOMETRIC MONITORING OF ALPHA-AMINO-ACIDS WITH COPPER WIRE AND TUBULAR ELECTRODES [J].
ALEXANDER, PW ;
MAITRA, C .
ANALYTICAL CHEMISTRY, 1981, 53 (11) :1590-1594
[2]   APPLICATION OF A COPPER TUBULAR ELECTRODE AS A POTENTIOMETRIC DETECTOR IN THE DETERMINATION OF AMINO-ACIDS BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
ALEXANDER, PW ;
HADDAD, PR ;
LOW, GKC ;
MAITRA, C .
JOURNAL OF CHROMATOGRAPHY, 1981, 209 (01) :29-39
[3]   KINETICS AND MECHANISM OF OXIDATION OF AMINES AND ALCOHOLS AT OXIDE-COVERED NICKEL, SILVER, COPPER, AND COBALT ELECTRODES [J].
FLEISCHM.M ;
PLETCHER, D ;
KORINEK, K .
JOURNAL OF THE CHEMICAL SOCIETY-PERKIN TRANSACTIONS 2, 1972, (10) :1396-&
[4]   HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY OF ORGANIC-ACIDS WITH POTENTIOMETRIC DETECTION USING A METALLIC COPPER ELECTRODE [J].
HADDAD, PR ;
ALEXANDER, PW ;
TROJANOWICZ, M .
JOURNAL OF CHROMATOGRAPHY, 1984, 315 (DEC) :261-270
[5]   ELECTROCHEMICAL DETERMINATION OF HISTAMINE DERIVATIZED WITH ORTHO-PHTHALALDEHYDE AND 2-MERCAPTOETHANOL [J].
HARSING, LG ;
NAGASHIMA, H ;
VIZI, ES ;
DUNCALF, D .
JOURNAL OF CHROMATOGRAPHY, 1986, 383 (01) :19-26
[6]   AMPEROMETRIC DETECTION OF AMINES AND AMINO-ACIDS IN FLOW-INJECTION SYSTEMS WITH A NICKEL-OXIDE ELECTRODE [J].
HUI, BS ;
HUBER, CO .
ANALYTICA CHIMICA ACTA, 1982, 134 (JAN) :211-218
[7]  
JACOBS W, 1986, CURRENT SEPARATIONS, V7, P39
[8]  
JI H, 1988, CHINESE J CHROMATOGR, V6, P1394
[9]  
KRAFIL JB, 1982, ANAL CHIM ACTA, V139, P347
[10]   CONSTANT POTENTIAL AMPEROMETRIC DETECTION AT A COPPER-BASED ELECTRODE - ELECTRODE FORMATION AND OPERATION [J].
LUO, PF ;
PRABHU, SV ;
BALDWIN, RP .
ANALYTICAL CHEMISTRY, 1990, 62 (07) :752-755